UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20201102

2020/11/02 (Monday)

Previous Day

  • Lab in-person: Ruiren, later Matt
  • Virtual partner: Matt, later Zach,Michael,Derek,Pramod
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → pi-filters 1/2/3 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)-master; DCRC33(2) → HV control. DCRCs on. Lead/poly shield erected (4 lead walls, 2 poly walls).
Status Doing a partial warmup because I suspect helium films on the detector are making the phonon channels behave funny. Suspended circulation yesterday. In the lab today we'll stop the 1K Pot. Mixture is still in fridge, V6 closed, V3 open, 3He pump on.
0800 MB = 29.3%
0900 Tower = 1.34 K, slowly rising. G1(fridge) = 33.1 mbar.
0900 Ruiren temperature = 36.6 C
1010 Ruiren enters lab
1015 Close 1K Pot Needle Valve. Close 1K Pot Speedivalve. Close valve at 4He Pump. Switch 4He pump outlet from Recovery to Exhaust.
1023 Tower/MC/Coldplate = 1.37 K/1.36 K/2.51 K, Still = 2.48 K / 1000 mbar / 0 mW, G1/G2= 34/27 mbar, 1K Pot = 2.35 K and rising, P2 = 0.015 mbar.
1023 transfer lN2 into cold trap dewar
1030 Ruiren leaves lab
1030 Close V4a. Turn off 4He pump.
1040 Tower/1K Pot = 1.39/3.0 K
1040 Turn on MC/Still heaters to 20/30 mW. Tower temperature starts rising fast.
1100 Tower = 3.15 K. As it rose there was a bit of a flattening around 2.0 K, and another longer one with some wiggles around 2.5 K. Maybe these are associated with the lambda point, as the helium on the tower warmed from superfluid to ordinary liquid.
Another temperature kink at 3.5 K, no theory for that one.
1130 Tower/1K Pot = 3.97/4.15 K. Tower temperature is flattening.
1130 Turn on DCRC1 LEDs with increased duty cycle: width/rate = 100/1000 us instead of standard 100/10000 us. Tower temperature rises more quickly.
1135 Tower = 4.13 K. It seems to have plateaued.
1145 G1 (fridge) = 906 mbar. Probably should let some into the dump soon.
1145 Turn off 3He pump. Fridge pressure falls a bit.
1155 G1(fridge) = 932 mbar. Open V4 for half a minute to let some mixture into dump. G1→800 mbar.
This has a cooling effect, Tower 4.2→4.05 K
1159 Tower suddenly starts rising from 4.2 K
1207 1K Pot suddenly starts rising from 4.5 K
1214 1K Pot thermometer maxes out at 7K
1230 Tower = 7.5 K. Turn off LEDs.
1230 G1(fridge) = 975 mbar. Open V4 for half a minute to let some mixture into dump. G1→773 mbar.
1235 Turn off MC/Still heaters.
1245 Tower has settled to 6.84 K
1315 Tower flat at 6.5 K
1420 Matt enters lab
1425 Matt Fritts body temperature = 97.5 F
1430 Start 4He pump, open pump valve, switch pump outlet back to Recovery
1440 Start LHe transfer @ MB = 19.3%
1445 Tower/1KPot = 6.80/>7 K
1445 Turn off DAQ programs, Unplug power from DCRC1&4
1445 Check pressure in 1K Pot by opening speedivalve (V4a closed): about 0.37 bar on mechanical gauge. 1K Pot temperature falls to 5.4 K
1455 Pump out 1K Pot in steps, alternating between speedivalve and V4a, opening briefly to keep pump inlet pressure low. 1K Pot → 3.6 K / 0.8 mbar.
1500 MC(H35)/Tower = 6.65/6.70 K
1500 Open 1K Pot needle valve → 10°. 1K Pot → 3.0 K / 7.5 mbar.
Open 1K Pot needle valve → 20°. 1K Pot → 2.6 K / 14 mbar.
G1 (fridge) is dropping, → 390 mbar. Tower falling slowly, → 6.62 K
1505 Open 1K Pot needle valve → 30°. 1K Pot → 2.4 K / 20 mbar. Oscillating. Tower temperature falling more quickly.
1510 Tower = 3.4 K and falling
1513 End LHe transfer @ MB = 91.2%
1525 Tower = 2.35 K and falling. 1K Pot = 1.54 K / 9.0 mbar.
Open 1K Pot needle valve → 40°. 1K Pot → 1.54 K / 9.0 mbar, no change
1535 Open 1K Pot needle valve → 50°. 1K Pot → 1.53 K / 8.5 mbar. Hmm, hope it isn't clogged.
1535 Tower = 2.17 K. G1(fridge) = 65 mbar. Turn on 3He pump.
1545 G1(fridge) = 64 mbar. Open V9. This lets the mixture from the dump into the fridge through cold LN2 Trap 1. G1(fridge)→166 mbar and rising. Tower temperature rises, 2.15→2.22 K. 1K Pot rises → 1.73 K / 190 mbar.
G1(fridge) peaked at 275 mbar then started falling. 1K Pot peaked at 200 mbar then started falling.
1550 Plug in DCRC1&4.
1550 Start DAQ programs and zap SQUIDs
1555 Matt leaves lab
1630 1K Pot = 50° / 1.65 K / 31±1 mbar. Tower = 2.00 K, G1 (fridge) = 54.1 mbar.
1645 Tower = 1.90 K, G1 (fridge) = 51.4 mbar.
1725 Tower = 1.85 K, G1 (fridge) = 46.6 mbar.
1725 Close V3 and V9, leaving 40 mbar (G2) in the dump.
1725 Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 243 mbar. (G2 peaks higher than G1.) Tower drops to 1.66 K.
1735 V6 → 13.2%. Condenser peaks at 297 mbar. Tower → 1.56 K
1750 V6 → 13.4%. Condenser peaks at 330 mbar. Tower → 1.56 K. Still → 700 mbar
1805 V6 → 13.6%. Condenser peaks at 414 mbar. Tower → 1.52 K. Still → 50 mbar.
1825 V6 → 14.0%. Condenser peaks at 514 mbar. Tower → 1.39 K. Still → 8.0 mbar.
1835 V6 → 14.4%. Condenser peaks at 504 mbar. Tower → 1.25 K. Still → 3.6 mbar.
1845 V6 → 15.0%. Condenser peaks at 436 mbar. Tower → 1.06 K. Still → 1.5 mbar.
There were some small wiggles as the Tower cooled below the 1.2 K Al transition temperature but mostly it seems to have plowed through it.
1855 V6 → 100.%. Condenser peaks at 458 mbar. Tower → 0.70 K. Still → 0.22 mbar.
1940 Tower = 99 mK, Still = 0.130 mbar, G1/G2 = 331/305 mbar.
1940 Open V14 to let the remaining mix out of the dump and into circulation. P1(Still) peaks at 55 mbar then falls again. Tower peaks at 99 mK. Condenser peaks at 421 mbar.
2030 Tower = 63.3 mK, Still = 0.380 mbar, G1/G2 = 310/285 mbar.
2030 Close V14