2020/11/01 (Sunday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Monica
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → pi-filters 1/2/3 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)-master; DCRC33(2) → HV control. DCRCs on (except DCRC4). Lead/poly shield erected (4 lead walls, 2 poly walls).
DCRC4 is powered off because it wasn't working right yesterday
0910 Not seeing pulses in DCRC2 channels, not sure what's up.
Zap DCRC1 SQUIDs
1045 Seeing pulses now
1055 Tower thermometer = 51.6 mK. DCRC1pD IbIs suggests detector temperature = 54.7 mK
Some ugly, spiky noise in DCRC1pD. Pulses still look like they have short falltimes, especially in DCRC1pB. Maybe the short fall times have to do with helium films rather than trapped flux?
1240 Matt Fritts body temperature 97.8 °F
1300 MB = 57.4%
1310 Matt arrives in lab
1310 Tower/MC/Coldplate = 51.0/29.9/188 mK, Still = 1.39 K / 0.110 mbar / 0 mW, G1/G2= 242/211 mbar, 1K Pot = 1.61 K / 17±1 mbar.
1315 Fill LN2 cold trap bath
1315 Plug in power on DCRC4
1315 Turn off DAQ programs, reenable DCRC4 on ODB, restart programs, zap DCRC1&4 SQUIDs.
1320 Turbo on IVC = 1.6 nbar. With Turbo valved out for 1 minute, IVC = nbar. Reopen Turbo to continue pumping IVC.
1320 Looks like DCRC4 channels are working, seeing pulses in them.
1705 Tower thermometer around 54 mK. DCRC1pD (Side1pF) IbIs implies detector temperature around 68 mK, quite high
1715 DCRC4pB IbIs is still behaving strangely, as though Rnormal was much higher than before.
1730 DCRC4pB (Side1pB) IbIs implies detector temperature around 57 mK (Tower thermometer also around 57 mK)
1745 DCRC4pC (Side1pD) IbIs implies it is completely in the normal state
1800 DCRC1pB (Side1pA) IbIs implies detector temperature probably around 68 or 69 mK
1820 DCRC1pC (Side1pC) IbIs implies detector temperature probably around 69 mK
so yeah - generally the TESs are behaving as if the detector is very warm (one channel is behaving oddly, but the others are generally consistent). Tower thermometer though reads pretty lo, so it's pretty strange. I'm guessing helium films. No point really taking data in this state. Should consider a partial warmup.
1830 Stop circulation to warm up - close V6. Still pressure starts to rise. After a few minutes G1/G2 pressures fall, Tower temperature rises. Still pressure rises.
1910 Tower = 0.71 K and rising, G1(condenser) = 64 mbar and falling, P1(still) = off scale.
1910 Open V3 to short Still and Condenser sides of fridge. G1(condenser) falls, → 31 mbar
2240 Tower = 1.0 K and rising slowly, G1(fridge) = 31 mbar and rising slowly.