UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20201008

2020/10/08 (Thursday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Michael, Ruiren / Pramod
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → pi-filters 1/2/3 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)-master; DCRC33(2) → HV control. DCRCs off. Lead/poly shield erected (4 lead walls, 2 poly walls).
Status Cooling after warmup to deal with clogged 1K Pot. Initial LHe transfer yesterday.
0915 MB = 65.3% / 174 mbar+1atm
0920 MC(Cernox)/Sorb/1K Pot = 5.10/8.4/6.58 K. G1(fridge) = 580.1 mbar (3He pump on).
0930 Matt Fritts body temperature = 97.4 F
1055 Matt enters lab
1100 Turbo on IVC = 2.5 nbar. With turbo valved out for 1 minute, IVC = 30 nbar. Open Turbo valve again, we'll keep pumping on IVC.
1100 Pump out line leading to 1K Pot with scroll pump. Then switch over to 4He pump. Turn on 4He pump then change pump exhaust to recovery and open valve to recovery lines.
1110 Check pressure in 1K Pot: Mechanical gauge = 0.24 bar. 1K Pot temperature dips to 4.8 K, then up to 5.7 K
1115 Pump out 1K Pot in steps, alternating between V4a and the 1K Pot speedivalve, opening V4a for short bursts only when the mechanical gauge read above a few inches. After 8 speedivalve iterations leave V4a and 1K Pot speedivalve open. 1K Pot → 3.5 K / 0.85 mbar. Sorb falls, MC falls a little.
1120 MC(H35)/Tower = 5.14/5.14 K. (Coldplate/Still = 2.515/2.657 kΩ = 19.7/6.7 K. These don't have good calibrations here.)
1125 MC(H35)/Sorb = 5.14/4.3 K. G1 (fridge) = 550 mbar and falling.
1125 Open 1K Pot needle valve → 10°. 1K Pot → 2.7 K / 12 mbar.
Open 1K Pot needle valve → 20°. 1K Pot → 2.3 K / 30 mbar, oscillating. MC(H35) → 3.4 K. G1(fridge) → 460 mbar.
1135 1K Pot = 2.2 K / 30 mbar. MC(H35) = 3.16 K, G1 (fridge) = 366 mbar, both falling.
1145 1K Pot = 1.7 K / 30 mbar. MC(H35) = 2.82 K, G1 (fridge) = 237 mbar, both falling.
1150 Fill LN2 cold trap bath
1150 1K Pot = 1.68 K / 30 mbar, not oscillating. MC(H35) = 2.61 K, G1 (fridge) = 168 mbar, both falling.
1150 Open LN2 Dewar vent valve enough for slight flow, connect to Ebox purge
1200 1K Pot = 1.61 K / 17 mbar, pressure falling. MC(H35) = 2.21 K, G1 (fridge) = 78 mbar, both falling.
1200 Open 1K Pot needle valve → 30°. 1K Pot → 1.61 K / 17 mbar, not much change
MC is getting close to the lambda point, 2.17 K. I think 1K Pot pressure usually jumps at that point - we'll wait a bit
1205 MC(H35)/Tower = 2.12/2.13 K. Coldplate/Still = 3.106/3.176 kΩ =? 4.08/2.63 K.
1210 1K Pot = 1.58 K / 10 mbar. MC(H35) = 2.10 K, G1 (fridge) = 61 mbar, both flattening.
1210 Open 1K Pot needle valve → 40°. 1K Pot → 1.60 K / 14 mbar
MC(H35) = 2.07 K … now 1K Pot pressure/temperature shoot up, → 1.8 K / off scale, then start swinging…
1215 1K Pot = 1.75 K / 400 mbar, oscillating. MC(H35) = 2.04 K, G1 (fridge) = 59.7 mbar.
1250 MC(H35)/Tower = 1.94/1.95 K. Coldplate/Still = 3.190/3.178 kΩ =? 3.61/2.62 K.
1300 1K Pot = 40° / 1.75 K / 350±50 mbar.
Pot is running a bit high, let's step back on the needle valve
1300 Adjust 1K Pot needle valve → 35°. 1K Pot → 1.71 K / 65±5 mbar
1310 Matt leaves lab
1355 1K Pot = 35° / 1.71 K / 65±5 mbar. MC(H35) = 1.87 K, G1 (fridge) = 49.0 mbar.
1355 Close V3 and V9, leaving 39 mbar (G2) in the dump. Turn on 3He pump.
1400 Matt Fritts body temperature = 97.4 F
1400 Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 164 mbar. (G2 peaks higher than G1.) MC falls to 1.63 K. 1K Pot spikes then settles.
1405 V6 → 13.2%. Condenser peaks at 164 mbar. MC(H35) → 1.54 K
1410 V6 → 13.4%. Condenser peaks at 188 mbar. MC(H35) → 1.55 K (went up a little)
1420 V6 → 13.6%. Condenser peaks at 225 mbar. MC(H35) → 1.50 K. Still → 75 mbar.
1425 V6 → 14.0%. Condenser peaks at 204 mbar. MC(H35) → 1.15 K. Still → 4.5 mbar.
1440 Al superconducting transition: MC falls to 1.13 K, rises to 1.26, falls to 1.09, rises to 1.24, falls to 1.13, rises to 1.24, falls to 1.16, rises to 1.26, then starts to fall steadily
1450 MC starts rising again - this looks unusual. Rises to 1.4 K - Still and Condenser pressures rising strongly…
1455 MC/Condenser/Still peaked at 1.45 K / 260 mbar / 150 mbar, all falling now
1500 V6 → 14.4%. Condenser → 183 mbar. MC(H35) → 0.93 K. Still → 4.0 mbar.
1605 V6 → 15.0%. Condenser peaks at 216 mbar. MC(H35) → 0.61 K. Still → 2.4 mbar.
1615 V6 → 100%. Condenser peaks at 232 mbar. MC(H35) → 0.31 K. Still → 0.28 mbar.
1630 MC(H35) = 58 mK, Still = 0.170 mbar, G1/G2 = 177/169 mbar.
1650 MC(H35) = 33.4 mK, Still = 0.140 mbar, G1/G2 = 138/132 mbar.
1650 Open V14 to let the remaining mix out of the dump and into circulation. P1(Still) peaks at 65 mbar then falls again. MC(H35) peaks at 80 mK. Condenser peaks at 206 mbar.
1715 MC(H35) = 35.5 mK, Still = 0.800 mbar, G1/G2 = 140/133 mbar.
1715 Close V14
1715 MB = 52.4%
1745 MC(H35) = 31.4 mK, Still = 0.130 mbar, G1/G2 = 144/138 mbar. 1K Pot = 1.70 K / 60 mbar
MC looks kind of flat. This is pretty similar to Sep 19 - we had mediocre base at that time, Tower ~ 65 mK. Not as good as Aug 20.

Next Day