| 1015 | Turbo on IVC = 14 nbar |
| 1027 | With turbo valved out, IVC = 1.4 ubar |
| 1032 | Squirted helium near stripline elbow, leak detector 26 → 103 pbar*l/s. |
| 1043 | Loosened the upper knurled knob. |
| 1045 | Squirted helium near stripline elbow, then tightened knurled knob, leak detector rose to 100,000 pbar*l/s while knob was being loosened. |
| 1046 | Valve in turbo pump to get leak signal low again. |
| 1048 | Closed turbo valve. |
| 1049 | Squirted helium near stripline elbow, leak detector 33 → 220 pbar*l/s. |
| 1130 | Stop leak checking, slowly letting nitrogen into IVC. |
| 1147 | IVC is at 1 atm. Make some adjustments to the stripline elbow. Redid two KF connections so that T-pipe aligns better with elbow, with elbow pushed down to lowest position (the two knurled knobs in contact). |
| 1159 | Pump out IVC with scroll pump. |
| 1210 | Pumping out IVC with turbo. |
| 1214 | Switch to leak detector, signal low again, around 30 pbarL/s. Tried squirting helium near stripline elbow again, did not see any leak. |
| 1238 | Turbo on IVC = 40 nbar. |
| 1245 | Matt leaves lab. |
| 1305 | Turbo on IVC = 25 nbar. |
| 1349 | Matt Fritts body temperature = 98.6 F |
| 1335 | Turbo on IVC = 22 nbar. |
| 1405 | Turbo on IVC = 19 nbar. |
| 1535 | Turbo on IVC = 16 nbar. |
| 1205 | Turbo on IVC = 14 nbar. |