| 0025 | Turbo on IVC = 74 nbar. |
| 0205 | Turbo on IVC = 63 nbar. |
| 0415 | Turbo on IVC = 54 nbar. |
| 0700 | Turbo on IVC = 46 nbar. |
| 0955 | Camera monitoring Turbo pump fell off its perch. |
| 1408 | Matt Fritts body temperature = 97.0 F |
| 1430 | Matt arrives in lab |
| 1445 | Turbo on IVC = 40 nbar; with turbo valved out for 10 min IVC = 2.2 ubar. |
| 1500 | Leak detector on IVC LD = 29 pbar*L/s |
| 1505 | Squirted helium near stripline tube, no change in leak rate. |
| | Compare to Feb 17: then, squirting at South side of joint between knurled knobs cause signal to rise from 15 to 45 pbarL/s. With “generous” squirting the signal rose to 200 pbarL/s on Feb 17. Clearly it is much improved. |
| | We will prepare to cryoclean the mixure. |
| 1528 | Check IGH pressures G1=45mbar, P1=4.5mbar, pumping them out with scroll pump. |
| 1541 | Check pressure in warm LN2 trap 1 = 1.2 mbar, pump out with scroll pump. |
| 1545 | Check pressure in warm LN2 trap 2 = 1.7 mbar, pump out with scroll pump. |
| 1600 | Cold trap dewar filled partially with LN2. |
| | Surprisingly our LN2 Dewar seems to still have plenty of LN2 in it. We must have filled it right before the hibernation began, but still impressive that it held for 2 months. |
| 1605 | Matt leaves lab. |
| 1645 | Matt Fritts body temperature = 98.0 F |