2020/02/12 (Wednesday)

Previous Day

0900 Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.5%
0905 Increase V6 → 12.7% to restart flow
0910 Increase V6 → 12.8% to restart flow
0915 Increase V6 → 12.9% to restart flow
0935 Increase V6 → 13.0% to restart flow
0950 Increase V6 → 13.1% to speed up flow
1000 Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 570-855 mbar. Pump mixture to dump.
Removed 5.3 mbar.
1020 Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 13.0%
1045 Increase V6 → 13.1% to speed up flow
1120 Stop cleaning through Trap 2 after 60 min. G2 flow pressure was 580-700 mbar. Pump mixture to dump.
Removed 2.4 mbar.
1140 Check contents of warm LN2 Trap 1: 5.3 mbar. Pump out with scroll pump.
1200 Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 13.0%
V6 is open too much, G2 hits 980 mbar. Close V6 and V14, start over…
1205 Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.5%
1215 Increase V6 → 13.0% to restart flow
again, V6 is open too much, G2 hits 980 mbar. Annoying. Close V6 and V14, start over…
1215 Restart cryocleaning, V6 = 12.6%
1225 Increase V6 → 13.0% to restart flow
1305 Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 570-785 mbar. Pump mixture to dump.
Removed 5.4 mbar.
1540 Check contents of warm LN2 Trap 1: 5.4 mbar. Pump out with scroll pump.
1545 Check contents of warm LN2 Trap 2: 2.4 mbar. Pump out with scroll pump.