2020/01/27 (Monday)

Previous Day

1500 Electrical checks on side coaxes in cleanroom. Pins labeled based on either detector (D) or tower (T) end of side coax. Used HP 4275A LCR meter to measure Cc.
these are CDMSII position-4 length side coaxes. We haven't used them in a while but we will want to in the next run so we can put the source mover in the first position and the detector in the second position.
Side Coax : ZB4-4
Pin 1 Pin 2 Component Value
T1 D1 PAret 20 Ω
T2 D2 PA 20 Ω
T3 D3 PBret 20 Ω
T4 D4 PB 20 Ω
T5 D5 PCret 20 Ω
T6 D6 PC 20 Ω
T7 D7 PDret 20 Ω
T8 D8 PD 20 Ω
T9 D9 LED1 20 Ω
T10 D10 LED2 20 Ω
T11 D11 QO bias 26 nS
T12 D11 QO Cc <0.1 nS
T12 D11 QO Cc 296 pF
T12 T13 QO FB 26 nS
T14 D12 QI bias 26 nS
T15 D12 QI Cc >30 nS
T15 D12 QI Cc 291 pF
T15 T16 QI FB 26 nS
all pins GND - <0.1 nS
Side Coax : ZB1-4-6
Pin 1 Pin 2 Component Value
T1 D1 PAret 20 Ω
T2 D2 PA 20 Ω
T3 D3 PBret 20 Ω
T4 D4 PB 20 Ω
T5 D5 PCret 20 Ω
T6 D6 PC 20 Ω
T7 D7 PDret 20 Ω
T8 D8 PD 20 Ω
T9 D9 LED1 20 Ω
T10 D10 LED2 20 Ω
T11 D11 QO bias 26 nS
T12 D11 QO Cc <0.1 nS
T12 D11 QO Cc 295 pF
T12 T13 QO FB 26 nS
T14 D12 QI bias 25 nS
T15 D12 QI Cc 9 kΩ
T15 D12 QI Cc 295 pF
T15 T16 QI FB 26 nS
all pins GND - <0.1 nS
Side Coax : ZB1-4-8
Pin 1 Pin 2 Component Value
T1 D1 PAret 15 Ω
T2 D2 PA 15 Ω
T3 D3 PBret 15 Ω
T4 D4 PB 15 Ω
T5 D5 PCret 15 Ω
T6 D6 PC 15 Ω
T7 D7 PDret 15 Ω
T8 D8 PD 15 Ω
T9 D9 LED1 15 Ω
T10 D10 LED2 15 Ω
T11 D11 QO bias 25 nS
T12 D11 QO Cc <0.1 nS
T12 D11 QO Cc 325 pF
T12 T13 QO FB 25 nS
T14 D12 QI bias 26 nS
T15 D12 QI Cc <0.1 nS
T15 D12 QI Cc 328 pF
T15 T16 QI FB 25 nS
all pins GND - <0.1 nS

All phonon and LED channels are working. The coupling capacitors on ZB4-4 and ZB1-4-6 both have the right capcitances (~300 pF) but are also shorted. Perhaps the capacitors' dielectric is partly broken down?