2019/10/31 (Thursday)
| Status: At 77K discovered IVC leak near bottom of IVC. Warming now to check indium seal. Working toward 2nd attempt at Run 71 cooldown. | |
| 0800 | MC(Cernox) = 230.4 K |
| 0900 | MC(Cernox) = 234.35 K the warming time constant (room temp = 296 K) appears to be 16.1 hours, which suggests MC will reach dew point (281 K) tomorrow 7:45 am. |
| 0935 | Check contents of warm LN2 Trap 2: 4.9 mbar. This was open to mixture during the LN2 transfer. Pump out trap with scroll pump. |
| 0945 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.6% |
| 0950 | Increase V6 → 12.7% to restart flow |
| 1000 | Increase V6 → 12.8% to speed up flow |
| 1000 | MC(Cernox) = 238.05 K time constant ~ 16.2 hours, suggesting MC will reach dew point tomorrow 7:50 am. |
| 1145 | Increase V6 → 13.0% to restart flow |
| 1100 | MC(Cernox) = 241.5 K time constant ~ 16.3 hours, suggesting MC will reach dew point tomorrow 8:01 am. |
| 1155 | Increase V6 → 13.1% to speed up flow |
| 1200 | MC(Cernox) = 244.75 K time constant ~ 16.3 hours, suggesting MC will reach dew point tomorrow 7:58 am. |
| 1240 | Stop cleaning through Trap 1 after 175 min. G2 flow pressure was 560-750 mbar. Pump mixture to dump. Removed 3.8 mbar. That's a lot lower - maybe circulating during LN2 transfer did a lot of cleaning too, and maybe tightening the o-rings helped. Or maybe it's an anomaly. |
| 1300 | MC(Cernox) = 247.85 K time constant ~ 16.0 hours, suggesting MC will reach dew point tomorrow 7:41 am. |
| 1400 | MC(Cernox) = 247.85 K time constant ~ 16.1 hours, suggesting MC will reach dew point tomorrow 7:46 am. |
| 1415 | Check warm LN2 Cold Trap 1: 3.8 mbar. Pump out Trap 1. |
| 1435 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.9% |
| 1450 | Increase V6 → 13.1% to restart flow |
| 1455 | Increase V6 → 13.2% to restart flow |
| 1510 | Increase V6 → 13.3% to speed up flow |
| 1500 | MC(Cernox) = 253.5 K time constant ~ 15.9 hours, suggesting MC will reach dew point tomorrow 7:36 am. |
| 1535 | Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 570-980 mbar. Pump mixture to dump. Removed 5.6 mbar. |
| 1705 | Check warm LN2 Cold Trap 1: 5.6 mbar. Pump out Trap 1. |