2019/07/17 (Wednesday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)
DCRC33(2) → 1kV power supply → HV input

Three walls of the lead/poly shield are up.
Fridge failure yesterday. Last night I closed circulation and opened the fridge to the dump, remotely. It looked like maybe a sudden boiloff of helium in the IVC. But….
0835 upon arrival, find that the pumps are off. Maybe the fridge failure was just a power interruption?
0835 Close 1K Pot speedivalve. Then reset fuse box - pumps restart.
0840 Tower/1K Pot/Sorb = 5.5/>7/9.5 K. MB = 28.7%
not sure what to make of the >4K temperatures. It may be that we now have a lot of helium gas in the IVC. Could be that that's a symptom of the fridge failure rather than a cause. Should pump it out. Not this will be “blind” as we have no Pfeiffer gauge.
0845 Connect scroll pump to pump manifold, pump out manifold up to IVC valves (which are closed)
0850 Switch to pumping with turbo pump
0855 Stop pumping on manifold, then open IVC to pump manifold. Also have the mechanical valve in line - watch that valve - no increase when I open the IVC. No reaction from the fridge temperatures.
0855 Close valve to mechanical gauge, thecdms:k100:run_diary:20190525n open valve to turbo pump. Now pumping IVC with turbo pump.
0910 Close valve to 1K Pot pump, then open speedivalve. P2 rises off scale. Close needle valve.
0915 Pump out 1K Pot in steps, then start pumping it continuously.
0920 Open 1K Pot needle valve → 5°. 1K Pot → 3.0 K / 7 mbar
Open 1K Pot needle valve → 10°. 1K Pot → 2.8 K / 15 mbar
Open 1K Pot needle valve → 15°. 1K Pot → 2.4 K / 26 mbar
0925 Open 1K Pot needle valve → 20°. 1K Pot → 1.9 K / 90 mbar. Tower temperature falling, 5.4→4.0 K
0930 Fill LN2 cold trap full.
0935 P2 is decreasing, 1K Pot = 1.7 K / 36 mbar, Tower = 2.7 K
1010 1K Pot pressure jumped to 85 mbar and is oscillating +/- 10 mbar. 1K Pot = 1.7 K, Tower = 2.6 K. G1/G2 = 202/191 mbar and have been falling.
1220 1K Pot = 1.7 K / 85 mbar, pressure is oscillating as normal. Looks like it's working nicely.
But Tower is not cooling, looks stuck at about 2.5 K. This is a bad sign - it may indicate there's still a lot of helium in the IVC.
1225 Switch to Cernox thermometer to better compare to initial Run70 cooldown.
MC(Cernox) = 2.48 K
Last time it flattened out too, but at a lower temperature, 2.1 K. We could go ahead and try circulation I guess.
G1/G2 = 151/143 mbar and pretty flat. Last time they flattened a lot lower at 61 mbar.
1400 Commence LHe tranfer. MB=18.5%. Using 100L dewar. He gas cylinder is empty!? When did that happen? Replace it with a new one.
1442 Cease LHe transfer. MB=91.3%
1457 Close V9, leaving 132 mbar in the dump.
Start circulation through Trap 1. V6 → 12.5%. No immediate changes.
1500 V6 → 12.8%. No changes, MC (Cernox) = 2.5 K, G1 = 145 mbar
1503 V6 → 13.0%. Condenser flat at 149 mbar. MC (Cernox) → 2.5 K.
1507 V6 → 13.2%. Condenser flat at 150 mbar. MC (Cernox) → 2.5 K.
1510 V6 → 13.4%. Condenser peaks at 189 mbar. MC (Cernox) → 2.4 K.
1515 V6 → 13.6%. Condenser peaks at 357 mbar. MC (Cernox) → 2.0 K.
1519 V6 → 13.8%. Condenser peaks at 450 mbar. MC (Cernox) → 1.6 K. Still → 500 mbar.
1526 V6 → 14.2%. Condenser peaks at 400 mbar. MC (Cernox) → 1.4 K. Still → 9.5 mbar.
1534 V6 → 14.6%. Condenser peaks at 350 mbar. MC (Cernox) → 1.3 K. Still → 5.0 mbar.
1545 V6 → 15.0%. Condenser peaks at 350 mbar. Tower → 1.3 K. Still → 4.0 mbar.
1551 V6 → 15.4%. Condenser peaks at 368 mbar, rising slowly. Tower → 900 mK. Still → 3.0 mbar.
Cooling is behaving a bit strangely here, the condenser pressure G1 continues to slowly rise as the tower temperature continues to fall. This has been ongoing for 30 minutes and doesn't show any signs of stopping. (1620) Tower → 900 mK. G1 - > 366 mbar
1600 Fill LN2 Cold Trap Bath Full
1625 V6 → 16.0%. Condenser peaks at 377 mbar. Tower → 800 mK. Still → 1.9 mbar.
1630 V6 → 16.4%. Condenser peaks at 379 mbar. Tower → 717 mK. Still → 1.4 mbar.
1643 V6 → 16.8%. Condenser peaks at 380 mbar. Tower → 650 mK. Still → 1.1 mbar.
1646 V6 → 100%. Condenser peaks at 386 mbar. Tower → 520 mK. Still → 0.32 mbar. 1K Pot = 1.7K / 90 mbar
1730 Tower = 81 mK, Still = 0.260 mbar, Condenser = 311 mbar, 1K Pot = 1.72 K / 110 mbar
1730 Open V14 to let the remaining mix out of the dump and into circulation. P1(Still) rises off scale, falls again after a minute. Condenser peaks at 480 mbar. Tower → 175 mK then falls. 1K Pot peaks at 1.79 K / off-scale mbar then settles.
1740 Tower = 143 mK, Still = 4.0 mbar, Condenser = 162 mbar, 1K Pot = 1.67 K / 40 mbar. Pot temporarily rises then dips, then settles to something normal.
1750 Tower = 85 mK, Still = 1.5 mbar, Condenser = 137 mbar, 1K Pot = 1.71 K / 80 mbar.
1800 Tower = 79 mK, Still = 0.8 mbar, Condenser = 133 mbar, 1K Pot = 1.71 K / 85 mbar.
Close V14.