Previous Day

Status: LHe transfer yesterday. Weird issues writing to chocula continue.
0844 MC(Cernox) = 4.7 K, Sorb = 8.4 K, turbo on IVC = 0.86 nbar, G1 (fridge) = 559 mbar, DIB4 pD SQUID = 311 Ω
0849 Quick electrical check looks good.
0852 Plug DIB3 into DCRC3 and power it up. Start DCRC programs.
0855 Try to use DCRC Control tools to zap DCRC3 SQUIDs, but keep getting 'problem in tools timeout' error. I know that this just creates a sequence called 'testzap.msl'. Looking at the existing version of it, I see it is the correct sequence to zap just DCRC3 SQUIDs since that's the last time we used it. Just run 'testzap.msl' sequence through MIDAS sequencer so we can get on with things.
0900 Just looking at PSDs on the pulse viewer, it looks like DIB3 PD (SQUID only) has very low, flat noise at around 2.48e-7 @ 1kHz (in some units? Not sure how the SQUID tuning GUI PSD is normalized). DIB3 PB has a flat spectrum at 3.45e-6 with R/L roll off as expected for Johnson noise.
0902 Let's try a baby IbIs with DCRC3 pB. Looks nice and linear with slope of -16.9 mV/uA
As described on the Run 70 summary, this implies a wirebond (pair) resistance of 5.0 mΩ
Ib (uA) Vo (V)
-10 -0.1753
-5 -0.2613
0 -0.3445
5 -0.4284
10 -0.5129
0906 Start 100s series with 20 Hz in run randoms and 1000 BOR randoms, DCRC3 PB/PD QET bias = 0, and phonon trace length = 52.4288 ms
0908 MC(Cernox) = 4.7 K, Sorb/1K Pot = 8.4/6.6 K, G1 (fridge) = 559 mbar
0909 Pump out lines leading to 1K Pot with scroll pump
0911 Switch over to 4He pump. Change pump exhaust to recovery and open valve to recovery lines.
0916 Check pressure in 1K Pot by closing V4A and opening 1K Pot speedivalve: the mechanical gauge goes up about 0.3 bar
This brought the 1K Pot temp down to 3.5 K
0918 Pump out 1K Pot in steps, alternating between V4a and the 1K Pot speedivalve. After 5 iterations get pot down to < 0.1 bar. Then leave 4He pump pumping on pot continuously.
0921 1K Pot = 3.2 K / 3.2 mbar. MC (Cernox) = 4.69 K and falling slowly, G1 = 533 mbar and falling slowly.
0925 1K Pot = 3.2 K / 3.2 mbar, steady.
0926 Open 1K Pot needle valve → 10°. 1K Pot → 2.7 K / 8.0 mbar
Needle valve was closed about 5° past the 0 marked on the brass knob. It was a little sticky to open at first, but not as bad as yesterday.
0927 MC = 4.6 K and falling, G1 = 484 mbar and falling.
0930 Open 1K Pot needle valve → 20°. 1K Pot → 2.4 K / 19.0 mbar
0932 Open 1K Pot needle valve → 30°. 1K Pot → 2.0 K / 38.0 mbar
0935 MC = 3.1 K and falling G1 = 375 mbar and falling.
0938 Open 1K Pot needle valve → 40°. 1K Pot pressure (P2) spikes to 260 mbar, but then starts falling slowly.
0940 1K Pot = 1.8 K (flat) / 190 mbar (falling), MC = 2.8 K and falling G1 = 248 mbar and falling.
0945 Do another baby IbIs. Linear with a slope of -17.1 mV/uA giving wirebond (pair) resistance of 4.8 mΩ
Ib (uA) Vo (V)
-10 -0.0959
-5 -0.1821
0 -0.2679
5 -0.3525
10 -0.4378
0947 1K Pot = 1.7 K / 55 mbar (falling), MC = 2.5 K and falling G1 = 135 mbar and falling.
0949 Just looking at PSDs on the pulse viewer, DIB3 PD (SQUID only) has very low, flat noise at around 2.36e-7 @ 1kHz. DIB3 PB has a flat spectrum at 2.40e-6 with R/L roll off.
0955 The chocula issues are messing with the fridge temp plots on the K100 status website. Temporarily move this code from chocula to my personal directory and make a link to it from the chocula location.
Ughh,that doesn't help because the logs are all supposed to be written to chocula. Not sure if the fridge logs are being written. I notified IT about this yesterday, but haven't heard back.
Looks like the data I took got copied to chocula fine though, so that's good.
1000 1K Pot = 1.6 K / 16 mbar (flattening out), MC = 2.1 K, G1 = 63 mbar. Everything is starting to flatten out.
1003 Spontaneous jump in 1K Pot pressure → 1.75 K / 300 mbar and oscillating strongly.
1006 Close 1K Pot needle valve → 35°. 1K Pot → 1.70 K / 150 mbar and oscillating
1008 MC = 2.1 K, G1 = 62 mbar. Both quite flat.
1013 Close 1K Pot needle valve → 30°. 1K Pot → 1.72 K / 85 mbar. That seems a little more stable.
1016 Fill LN2 cold trap bath.
1017 Close V3 and V9, leaving 61 mbar in the dump.
Start circulation through Trap 1. V6 → 12.5%. No immediate changes.
1020 V6 → 12.8%. No changes, MC (Cernox) = 2.0 K, G1 = 58 mbar
1024 V6 → 13.0%. MC (Cernox) = 2.0 K, G1 = 60 mbar, 1K Pot = 1.72 K / 75 mbar
1026 V6 → 13.2%. Condenser peaks at 82 mbar. MC (Cernox) falls to 1.97 K. 1K Pot steady at 1.72 K / 70 mbar
1029 V6 → 13.4%. Condenser peaks at 138 mbar. MC (Cernox) falls to 1.56 K. 1K Pot steady at 1.72 K / 70 mbar
1030 Let's leave 10uA bias going down DCRC3pB and just watch the DC level. I think this should go down when the wirebonds go SC. DCRC3pB = -0.4375 V
1034 V6 → 13.6%. Condenser peaks at 130 mbar. MC (Cernox) falls to 1.47 K. 1K Pot steady at 1.72 K / 70 mbar. P1 comes within range
1038 V6 → 13.8%. Condenser peaks at 125 mbar. Still → 13 mbar. Tower → 1.28 K. DCRC3pB = -0.4373 V
Switched over to tower thermometer
1042 V6 → 14.0%. Condenser peaks at 120 mbar. Still → 4 mbar. Tower → 1.20 K. DCRC3pB = -0.4379 V
1044 V6 → 14.4%. Condenser → 132 mbar. Still → 2.6 mbar. Tower → 0.99 K. DCRC3pB = -0.4381 V
1050 V6 → 15.0%. Condenser → 136 mbar. Still → 1.3 mbar. Tower → 874 mK. DCRC3pB = -0.4383 V
1052 V6 → 16.0%. Condenser → 136 mbar. Still → 0.60 mbar. Tower → 717 mK. DCRC3pB = -0.4381 V
1052 V6 → 100%. Condenser → 137 mbar. Still → 0.20 mbar. Tower → 591 mK. DCRC3pB drops to -0.4428 V, then starts rising again.
1K Pot = 1.72 K / 80 mbar
1101 Omni-logger saves it's log to chocula. Today's log looks screwy, it's missing the header and hasn't been updated in a while. Probably due to chocula write issues. Stop OMNI-logger, delete today's temp log file and restart Omni-logger. Now the log file looks correct and is updating.
1105 Tower = 400 mK, Still = 0.170 mbar, Condenser = 130 mbar, 1K Pot = 1.72 K / 75 mbar. DCRC3pB = -0.4381 V
1110 Fix a badly formatted line in yesterday's log too. Now the K100_Plotter script should work and keep the website updated.
1115 Tower = 229 mK, Still = 0.140 mbar, Condenser = 108 mbar, 1K Pot = 1.71 K / 70 mbar. DCRC3pB = -0.4378 V
1124 Tower = 88.6 mK, Still = 0.110 mbar, Condenser = 90 mbar, 1K Pot = 1.72 K / 80 mbar. DCRC3pB = -0.4379 V
1128 Open V14 to let the remaining mix out of the dump and into circulation. P1(Still) rises off scale, falls again after a minute. Condenser peaks at 115 mbar. Tower → 100 mK and rising. 1K Pot peaks at 1.72 K / 100 mbar then settles.
1133 Tower peaks at = 128 mK, Still = 6.0 mbar, Condenser = 79 mbar, 1K Pot = 1.72 K / 70 mbar. DCRC3pB = -0.4379 V
1140 MB = 54.4% Depending on the boiloff rate, this could be down to 20% by as early as 8AM tomorrow. Let's do a little fill now to get on a schedule of non-ish transfers.
1141 Tower = 94.3 mK, Still = 2.20 mbar, Condenser = 78 mbar, 1K Pot = 1.72 K / 80 mbar. DCRC3pB = -0.4378 V
1150 Modify the webcamAutomater.py script to accept device number as a parameter. Plug recovery webcam into jiajin.
Run python /home/mast/cdms/webcam webcamAutomater.py -device /dev/video1 -delay_min 0.033 -total_time_hr 3 -save_images to save images every ~2 sec
1152 P1 = 0.85 mbar, close V14.
1155 Commence LHe transfer, MB = 53.4%
1K Pot pressure runs off scale, then settles back down and starts to oscillate around 150 mbar
1200 Tower = 83.4 mK, Still = 0.110 mbar, Condenser = 87 mbar, 1K Pot = 1.73 K / 160 mbar (oscillating). DCRC3pB = -0.4387 V
1210 Tower = 81.9 mK, Still = 0.110 mbar, Condenser = 84 mbar, 1K Pot = 1.71 K / 80 mbar . DCRC3pB = -0.4375 V
1211 Turn on heater on LHe dewar to 4PSIg
1213 Temperature log has stopped updating again. :/ Ohh, I see now that chocula is full. That's a problem.
1235 Cease LHe transfer, MB = 91.3%
1237 Tower/MC/Coldplate = 81.7/33.5/200 mK, Still = 1.0±0.1 K / 0.110 mbar, G1/G2 = 83/92 mbar, 1K Pot = 1.68 K / 45 mbar. MB = 91.3%. DCRC3pB = -0.4369 V
Still is usually at ~1.7 K
1251 Fill LN2 cold trap.
1300 Things look mostly stable, though P2 falling slowly, hopefully it settles out. The temperature logging isn't working, but we should still be able to VNC to vuk1 to check on things.
1430 Tower = 80.7 mK, 1K Pot = 1.7 K / 36 mbar
1630 Tower = 79.9 mK, 1K Pot = 1.66 K / 32 mbar
2150 Tower = 78.2 mK, 1K Pot = 1.65 K / 30 mbar

Next Day