2019/06/12 (Wednesday)
| Setup | DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1) DCRC33(2) → 1kV power supply → HV input Three walls of the lead/poly shield are up. |
|---|---|
| 0920 | Tower/MC/Coldplate = 42.3/30.8/199 mK, Still = 1.275 K / 0.090 mbar, G1/G2 = 153/144 mbar, 1K Pot = 1.63 K / 22 mbar. MB = 38.0% after several days in their “low” modes, Coldplate and Still thermometers are back in their “high” modes. |
| 0925 | Fill LN2 Cold trap bath |
| Last night's data was taken with a HV drift rate of +2.34 V/min, but it looks like that's still a little low, by about 0.08 V/min. Let's adjust to 2.418 V/min = 0.0403 V/s. Let's make sequences for 180V, 200V, 220V. |
[fritts@jiajin /home/hep/cdmssoft/scdmsDAQ/MidasDAQ/online/scripts/pysequencer]: python HVscan.py -iDCRC 2 -HVs 200 -HVcalFile HV_calibrations/HV_DCRC2_control_cals_440V_190603.txt -HVdriftRate 0.0403 -HVpreBias 10/2 -tSeries 82 -DCRCs2Flash 1,4 > ../../sequencer/HVdrift_200V_2p418VPM_prebias_10_2_dur_82min_flashDCRC_1_4.msl [fritts@jiajin /home/hep/cdmssoft/scdmsDAQ/MidasDAQ/online/scripts/pysequencer]: python HVscan.py -iDCRC 2 -HVs 180 -HVcalFile HV_calibrations/HV_DCRC2_control_cals_440V_190603.txt -HVdriftRate 0.0403 -HVpreBias 10/2 -tSeries 90 -DCRCs2Flash 1,4 > ../../sequencer/HVdrift_180V_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl [fritts@jiajin /home/hep/cdmssoft/scdmsDAQ/MidasDAQ/online/scripts/pysequencer]: python HVscan.py -iDCRC 2 -HVs 220 -HVcalFile HV_calibrations/HV_DCRC2_control_cals_440V_190603.txt -HVdriftRate 0.0403 -HVpreBias 10/2 -tSeries 74 -DCRCs2Flash 1,4 > ../../sequencer/HVdrift_180V_2p418VPM_prebias_10_2_dur_74min_flashDCRC_1_4.msl
| 1030 | Let's try a negative bias. Reverse the leads on the HV supply. |
|---|---|
| 1035 | Notice that the HV is not very close to zero. Adjust DCRC2qI from 11.537 to 11.58 and find it's much closer to 0V (-1.5V → -0.011V). |
| 1050 | Create a new HV calibration file HV_calibrations/HV_DCRC2_control_cals_440V_190612.txt |
| 1055 | Create a new sequence for 200V+2.481V/min with the new calibration file: |
[fritts@jiajin /home/hep/cdmssoft/scdmsDAQ/MidasDAQ/online/scripts/pysequencer]: python HVscan.py -iDCRC 2 -HVs 200 -HVcalFile HV_calibrations/HV_DCRC2_control_cals_440V_190612.txt -HVdriftRate 0.0403 -HVpreBias 10/2 -tSeries 82 -DCRCs2Flash 1,4 > ../../sequencer/HVdrift_200V_cal0612_2p418VPM_prebias_10_2_dur_82min_flashDCRC_1_4.msl
| 1105 | Start an 82 min series at HV = -200→-400V, rate -2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. Use sequence HVdrift_200V_cal0612_2p418VPM_prebias_10_2_dur_82min_flashDCRC_1_4.msl |
|---|---|
| 1232 | Looks like the trigger rate is much too high, so increase DCRC1pB/C/D trigger thresholds to 8/8/8. Unfortunately this is pretty close to the end of the data series. |
| At the end of the sequence adjust DCRC2qI to get HV closer to zero, it was around 1V | |
| The HV drift doesn't look fast enough for this -200V data, for some reason. Should be 0.0063 V/s faster, so let's try 2.796 V/min = 0.0466 V/s. Surprising that it needs to be so high. |
[fritts@jiajin /home/hep/cdmssoft/scdmsDAQ/MidasDAQ/online/scripts/pysequencer]: python HVscan.py -iDCRC 2 -HVs 200 -HVcalFile HV_calibrations/HV_DCRC2_control_cals_440V_190612.txt -HVdriftRate 0.0466 -HVpreBias 10/2 -tSeries 71 -DCRCs2Flash 1,4 > ../../sequencer/HVdrift_200V_cal0612_2p796VPM_prebias_10_2_dur_71min_flashDCRC_1_4.msl
| 1344 | Start a 71 min series at HV = -200→-400V, rate -2.796V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 8/8/8. Use sequence HVdrift_200V_cal0612_2p796VPM_prebias_10_2_dur_71min_flashDCRC_1_4.msl |
|---|
- Noise width for this -200V series is around 38 eVee, significantly worse than for the +200V case
| 1420 | Begin LHe transfer. MB = 30.8% |
|---|---|
| 1500 | End LHe transfer. MB = 89.3% |
| 1550 | Fill LN2 Cold trap bath |
| 1630 | Switch HV leads back to positive polarity. Also, the HV calibration seems to have shifted a little - adjust the HV_DCRC2_control_cals_440V_190612.txt a little, just the near-zero values - the higher biases don't seem to have changed significantly. |
| We'll go back to the 2.418 V/min drift, which seemed correct for positive-bias data, even though for some reason we needed a faster drift for negative-bias. And we'll try the lower thresholds again. | |
| 1647 | Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl |
| 2022 | Start a 74 min series at HV = 220→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. Use sequence HVdrift_220V_cal0612_2p418VPM_prebias_10_2_dur_74min_flashDCRC_1_4.msl |
| 2245 | Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl |