2019/06/13 (Thursday)
| Setup | DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1) DCRC33(2) → 1kV power supply → HV input Three walls of the lead/poly shield are up. |
|---|---|
| 0715 | Start a 74 min series at HV = 220→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. Use sequence HVdrift_220V_cal0612_2p418VPM_prebias_10_2_dur_74min_flashDCRC_1_4.msl |
- Noise widths based on PTOF (unbalanced) for +180V, +200V, +220V → 400V data in the last couple days:
- 180V: 20.6 eVee
- 200V: 24.2 eVee
- 220V: 26.6 eVee
| 0905 | Fill LN2 Cold trap bath |
|---|---|
| 0915 | Tower/MC/Coldplate = 42.1/32.8/201 mK, Still = 1.29 K / 0.090 mbar, G1/G2 = 148/136 mbar, 1K Pot = 1.64 K / 24 mbar. MB = 67.7% italicized values were taken while the tower was still cooling after a flash. |
| 1054 | Start a 99 min series at HV = 160→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. Use sequence HVdrift_160V_cal0612_2p418VPM_prebias_10_2_dur_99min_flashDCRC_1_4.msl |
| 1402 | Start a 100 min series at HV = 140→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait. CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. Use sequence HVdrift_140V_cal0612_2p418VPM_prebias_10_2_dur_100min_flashDCRC_1_4.msl |
- Noise width based on PTOF (unbalanced):
- 160V: 22.9 eVee
- 140V: 25.1 eVee
| 1645 | Fill LN2 Cold trap bath |
|---|---|
| 1650 | HV is meant to be set at 0, but is reading 2.2 V. Why is that calibration drifting now? Adjust the low end of the calibration file HV_calibrations/HV_DCRC2_control_cals_440V_190612.txt |
| 1711 | Start a 30 min series at 0V CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5) 1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6. |