2019/03/27 (Wednesday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)
DCRC33(2) → 1kV power supply → HV input


Lead poly shield is erected, N/E/W walls complete, no poly or lead on S wall.

DCRCs 3&4 powered down, readout cables disconnected at Ebox
1000 Tower/MC/Coldplate = 39.4/29.2/184 mK, Still = 1.65±0.10 K / 0.150 mbar, G1/G2 = 269/254 mbar, 1K Pot = 1.78 K / 14 mbar. MB = 68.7%
1015 Fill LN2 cold trap bath
1430 Remove bias resistors connecting unused side 2 phonon lines to HV on HV board. This leaves only the DIB4 PA phonon lines connected, of which the QET and QET ret lines are connected to the contact-free electrode. The full description of the current HV board status can be found on the hv board page.
1435 Tower = 39.5 mK
1440 Plug DCRCs back into HV board. All DCRCs are off right now. Plug detector cables into HV board and into ebox. No real temp change.
1445 Power on DCRCs 2,3,4. Enable DCRCs 2,3,4 in ODB. Try to start DCRC driver, but it won't start. Can telnet to all DCRCs without issue. It finally just started up after hitting the button a few times…
1450 Tower = 41 mK and rising.
1452 MIDAS was still being weird. cdms2 is really sluggish. Restart MIDAS and dcrc programs. Tower = 46.3 mK and rising.
1456 Turn on and off 15V power on DCRCs 3,4
1458 Tower temp = 55 mK and still rising. Stop DCRC programs and power down DCRCs 3,4.
1501 Tower temp peaks at 57.5 mK and starts to fall again.
1555 MB = 60.5%
1555 Tower = 39.6 mK, back down to this morning's temperature.
1600 Try powering DCRC4 only. First set up the ODB to enable DCRC4 & 2 only, and set DCRC4 as timing master. Power the DCRC then turn on the DAQ programs.
1605 Tower = 43.5 mK and climbing. Looks like it's rising as fast as before. Maybe it will slow down…
1605 Looking at DCRC4 phonon channels. DCRC4pA seems to not respond to QET bias changes. Try zapping SQUIDs.
1610 Tower peaked at 47.2 mK but then started falling. This happened while I was trying to get the phonon channels back in shape - don't know whether it's related.
1620 Messed around with phonon channels, after a while got all three DCRC4 channels in transition. During the process, temperature started falling more quickly. Looks like I “fixed” something.
1625 Tower = 45.0 mK and falling
1625 Reset MIDAS for DCRC3&4, power up DCRC3, restart DAQ programs. Tower temperature still falling.
1625 Zap DCRC3 SQUIDs.
1650 Got all channels into transition except DCRC3pB, which seems to be semi-unresponsive.
Tower temperature falling, now 42.4 mK
1657 Fill LN2 cold trap full.
1705 DRCC3 pB mostly unresponsive. Try zapping its SQUID.
1710 Tower = 40.3 mK and falling.
1715 DCRC3 pB now responds a little bit to changing from open loop/non inv. Also small offset changes with QET bias, but no pulses.