2019/03/18 (Monday)
| Setup | DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3) DCRC33(2) → 1kV power supply → HV input note: QET biases left at 0 overnight |
|---|---|
| 0910 | Tower/MC/Coldplate = 44.1/30.0/191 mK, Still = 1.70±0.20 K / 0.150 mbar, G1/G2 = 228/202 mbar, 1K Pot = 1.78 K / 16 mbar. MB = 42.8% |
| Tower falling 1.1 mK/day | |
| 0920 | Fill LN2 cold trap bath |
| 1058 | Start superconducting background series at HV = 0V CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = 0 uA, Front-End Gains = (4/2/3)/(10/15/10), trace length=65536 samples 1000 BOR randoms, then 20 Hz randoms |
| 1103 | Stop superconducting data |
| 1114 | Start normal-state background series at HV = 0V CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = 800 uA, Front-End Gains = (4/2/3)/(10/15/10), trace length=65536 samples 1000 BOR randoms, then 20 Hz randoms |
| 1121 | Stop normal-state data |
| 1135 | Get all 6 channels back into reasonable transition positions. |
| 1145 | Estimate event rate from looking at 30 52.488 ms traces in DCRC3pB, count a total of 155 pulses, so rate is 99±8 Hz. |
| 1205 | Install W and E walls of lead/poly shield |
| 1230 | Add 5 layers of 1/8“ thick lead sheets under cryostat. These fit between the posts with corner cut-outs for the posts. 22” N-S by 26“ E-W with 2”x2“ cutouts at corners (or a little bigger to accommodate the posts). |
| 1255 | Add |
| Note there are also permanent layers on the platforms, E and W sides, 10“ by 41” | |
| 1355 | Estimate event rate from looking at 30 52.488 ms traces in DCRC3pC, count a total of 117 pulses, so rate is 74±7 Hz. |
| 1440 | Finish poly on North shield wall |
| 1555 | Begin LHe transfer @ MB = 33.2% |
| 1600 | Isolate 4He pump exhaust from recovery, so we can look for any change in between-transfer recovery efficiency. |
| 1630 | End LHe transfer @ MB = 92.1% |
| 1650 | Fill LN2 Cold Trap bath |