2018/08/16 (Thursday)
| Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3) 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4) DCRC33(2) → kepco 1kV power supply → HV input |
|
| status: Took SiHV+NaI PuBe data overnight. | |
| 0910 | Tower/MC/Coldplate = 31.4/29.3/164 mK, Still = 1.0±0.1 K / 0.080 mbar, G1/G2 = 137/130 mbar, 1K Pot = 1.64 K / 30 mbar, Turbo on IVC = 1.0 nbar, MB = 28.4% |
| 0940 | Fill LN2 cold trap bath. |
| 0940 | Place a x10 attenuator on each of the NaI amplifier outputs. Setup: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → x10/x10 attenuator → DCRCX(4) |
| 0945 | Remove 22Na source from safe and place on top of southwest corner of shield. |
| 0954 | Start 1hr NaI calibration dataset HV=0V, with CDMS and NaI detectors running and DCRC2 as HV controller. CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10 NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC4pA/B threshold = 2/2 |
| 1101 | Start 1hr NaI calibration dataset HV=0V, with CDMS and NaI detectors running and DCRC2 as HV controller. CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10 NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC4pA/B threshold = 2/2 |
| 1215 | Return 22Na source to safe. |
| 1216 | Start 1 x 1hr 40 V background datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×1-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10 NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9 |
| 1335 | Begin LHe transfer @ MB = 20.4% |
| 1507 | Cease LHe transfer @ MB = 91.2% |
| 1550 | Start 6 x 2hr 40 V background datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×6-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10 NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9 MIDAS crashed during the 5th series |
| 1655 | Fill LN2 cold trap bath |