2018/08/17 (Friday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: Took SiHV+NaI PuBe data overnight. MIDAS crashed at some point around 0341 during the 5th series.
1010 Fill LN2 cold trap bath.
1010 Restart MIDAS, dial down HV, flash for 30 sec
1030 Tower/MC/Coldplate = 31.2/29.8/182 mK, Still = 1.4±0.1 K / 0.075 mbar, G1/G2 = 122/114 mbar, 1K Pot = 1.62 K / 24 mbar, Turbo on IVC = 1.2 nbar, MB = 67.3%
1125 SQUIDTuningGUI failing to start with error: TMidasControl::connect: Connecting to experiment “cdms_UMN” on host “cdms2.spa.umn.edu:7071” TMidasControl::connect: Cannot connect to MIDAS, status 503. Restarting MIDAS doesn't help. Restarting the terminal in the vnc session doesn't help. Restarting the vnc session itself did the trick. Not sure why, not gonna think too hard about it.
1148 Start 1 x 2hr 40 V background datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×6-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1307 Begin LHe transfer @ MB = 62.0%
1439 End LHe transfer @ MB = 94.2%
1508 Start 10 x 2hr 40 V background datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×10-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1700 Fill LN2 cold trap full.