| | These look reasonable. Due to the low shunt resistance, we can't know if a single bind is popped from this measurement. It at least indicates that no channel had both wirebonds popped. |
| 1115 | IVC = 8.6 μbar with scroll pump pumping on it. Switch to pumping with turbo pump |
| 1251 | Take LN2 trap 1 out to warm. |
| 1246 | Open V1 to look at He cold trap contents. G1 = 1.3 → 7.5 mbar. So 6.2 mbar of contaminants. Pump this out using the 1K Pot pump set to exhaust. |
| 1310 | Tower = 217 K. Perform full electrical check. Situation normal. |
| 1320 | Put Leak Checker on IVC. Background leak rate = 640 pbarL/s. Compare 55 pbarL/s earlier in this run. Of course the main bath is full of He now though. |
| 1354 | Pump out main bath with scroll pump. |
| 1400 | P2 < 1 mbar while pumping, leak rate = 600 pbarL/s |
| 1405 | Flush main bath with Nitrogen He. Forgot to change over to N2 line. Leak rate rises to 1.47 nbarL/s when we do this. |
| 1406 | Pump out main bath with scroll pump. |
| 1410 | P2 < 1 mbar while pumping, leak rate = 610 pbarL/s and flattening out. |
| 1420 | Flush main bath with NITROGEN. Leak rate = 570 pbarL/s. I don't think we will be able to see any indium leaks with this background. |
| 1424 | Spray test warm IVC plumbing on top of fridge. Background leak rate = 560 pbarL/s. No increase seen. This background may be too high to see anything.
Switch back to turbo pump. Let's pump on the IVC all weekend and hope this background comes down. |
| 1445 | With turbo valved out, IVC = 46 ubar. Resume turbo pumping. |
| 1446 | Tower = 224 K |