UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20180511

2018/05/11 (Friday)

Previous Day

Status: Opened IVC yesterday to look for touch from cold stage to 4K. Pumping out IVC overnight. Cryostat vacuum jacket was pumped out for a few hours
0900 Turbo on IVC = 16 nbar, IVC = 340 nbar with pump valved out
0945 Leak detector on IVC = 55 pbarL/s. Squirt-tests at IVC seal, no leak seen. At KF-flange on stripline elbow, 55→80 pbarL/s. Around knurled elbow flange, 60→90 pbarL/s.
1320 Raise main bath.
1330 Turbo on IVC = 16 nbar. IVC = 300 nbar with turbo valved out for a few minutes
1345 Pump out fridge lines with scroll pump
1345 Leak detector on IVC = 44 pbarL/s
1350 Let mixture from 3He pump through cold LN2 Trap 1 into fridge lines, both sides. (Most mixture still in dump.)
Leak signal does not change.
1350 Pump out 1K Pot with scroll pump. (Pumped it out earlier, had 26 mbar in it)
1355 Pump out main bath with scroll pump to <1 mbar
1405 Pressurize main bath with helium to 320 mbar above 1atm on G3. Do not let He into 1K Pot like last time. No increase in IVC leak.
1405 Let 5.4 mbar He exchange gas into IVC
1410 MC(Cernox) = 296.8 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 213 mbar, Old Level resistor = 0.985 kΩ, New level resistor = 2.142 kΩ
1415 Begin LN2 transfer
1420 MC(Cernox) = 296.2 K, IVC = 5.4 mbar, G1 (fridge) = 212 mbar, Old Level resistor = 1.058 kΩ, New level resistor = 2.162 kΩ
1425 MC(Cernox) = 292.8 K, IVC = 4.6 mbar, G1 (fridge) = 211 mbar, Old Level resistor = 1.086 kΩ, New level resistor = 2.187 kΩ
1430 MC(Cernox) = 286.2 K, IVC = 4.6 mbar, G1 (fridge) = 211 mbar, Old Level resistor = 1.109 kΩ, New level resistor = 2.205 kΩ
1442 MC(Cernox) = 268.9 K, IVC = 4.0 mbar, G1 (fridge) = 209 mbar, Old Level resistor = 1.154 kΩ, New level resistor = 2.300 kΩ
1455 MC(Cernox) = 243.1 K, IVC = ?? mbar, G1 (fridge) = 208 mbar, Old Level resistor = 1.240 kΩ, New level resistor = 2.858 kΩ
1500 MC(Cernox) = 229.5 K, IVC = 2.5 mbar, G1 (fridge) = 207 mbar, Old Level resistor = 1.262 kΩ, New level resistor = 2.885 kΩ
1506 MC(Cernox) = 213.7 K, IVC = 2.5 mbar, G1 (fridge) = 207 mbar, Old Level resistor = 1.288 kΩ, New level resistor = 2.915 kΩ
1516 MC(Cernox) = 191.4 K, IVC = 2.2 mbar, G1 (fridge) = 206 mbar, Old Level resistor = 1.310 kΩ, New level resistor = 2.958 kΩ
1526 MC(Cernox)/Sorb = 173.4/214.4 K, IVC = 2.2 mbar, G1 (fridge) = 205 mbar, Old Level resistor = 1.322 kΩ, New level resistor = 2.989 kΩ
1533 MC(Cernox)/Sorb = 163.3/194.4 K, IVC = 2.2 mbar, G1 (fridge) = 204 mbar, Old Level resistor = 1.327 kΩ, New level resistor = 3.000 kΩ
1541 MC(Cernox)/Sorb = 153.4/175.4 K, IVC = 2.2 mbar, G1 (fridge) = 204 mbar, Old Level resistor = 1.330 kΩ, New level resistor = 3.009 kΩ
1549 MC(Cernox)/Sorb = 145.0/160.1 K, IVC = 1.9 mbar, G1 (fridge) = 203 mbar, Old Level resistor = 1.334 kΩ, New level resistor = 3.015 kΩ
1557 MC(Cernox)/Sorb = 137.3/147.9 K, IVC = 1.9 mbar, G1 (fridge) = 202 mbar, Old Level resistor = 1.334 kΩ, New level resistor = 3.020 kΩ
1608 MC(Cernox)/Sorb = 128.1/136.6 K, IVC = 1.9 mbar, G1 (fridge) = 202 mbar, Old Level resistor = 1.336 kΩ, New level resistor = 3.021 kΩ
1615 MC(Cernox)/Sorb = 124.0/132.9 K, IVC = 1.9 mbar, G1 (fridge) = 202 mbar, Old Level resistor = 1.336 kΩ, New level resistor = 3.021 kΩ
1615 End LN2 transfer.
Cryocleaning mixture
0915 Begin cryocleaning mixture through cold LN2 Trap 2, V6 = 12.4%
0930 Increase V6→12.8% to speed up circulation
0940 Increase V6→12.9% to speed up circulation
0950 Increase V6→13.0% to speed up circulation
1015 Stop cryocleaning, Trap 2 60 minutes
1030 Begin cryocleaning mixture through cold LN2 Trap 1, V6 = 13.0%
1030 Increase V6→13.1% to speed up circulation
1040 Increase V6→13.2% to speed up circulation
1105 Increase V6→13.3% to speed up circulation
1135 Stop cryocleaning, Trap 1 65 minutes
1155 Check pressure in warm Trap 2: 9.2 mbar. Pump out with scroll pump.
1210 Begin cryocleaning mixture through cold LN2 Trap 2, V6 = 12.9%
1230 Increase V6→13.1% to speed up circulation
1240 Increase V6→13.3% to speed up circulation
1325 Stop cryocleaning, Trap 2 75 minutes. But realized v9 was open the whole time. So not really circulating properly.
1335 Check pressure in warm Trap 1: 17.6 mbar. Pump out with scroll pump.

Next Day