2018/07/23 (Monday)
| overnight: Tower temp, circulation pressures, 1K Pot mostly flat | |
| 0805 | Tower = 94 K, G1/G2 = 457/417 mbar, Still = 0.055 mbar / 0.1 mW, 1K Pot = 1.45 K / 4.0 mbar. |
| 0810 | Open 1K Pot needle valve 140→150°, to see what happens. Tower falls a little, Condenser rises a little. |
| 0815 | Turn off MIDAS DAQ programs, ground HV supply, power down all three DCRCs. (this causes a small spike in temperature) |
| 0815 | 1K Pot = 1.61 K / 20 mbar |
| 0820 | Tower = 86 mK, Condenser = 483 mbar. No improvement. We should warm up again and see if there's as much pressure in the IVC as before. |
| 0820 | Turbo on IVC = 2.5 nbar. Close valve to turbo pump. With pump valved out, IVC = 25 nbar. |
| 0923 | Fill LN2 cold trap full. |
| filling trap caused fall in Condenser 480-455 mbar, rise in Tower temperature 89→103 mK | |
| 1050 | Turn Still heater on to 5 mW, Still→0.110 mbar. Tower falls to 90 mK, at least initially. Condenser pressure rises above 500 mbar. |
| 1235 | Tower temperature started increasing again, now Tower = 120 mK, Condenser = 711 mbar. |
| 1240 | Suspend circulation: close V6, open V9 |
| 1310 | Tower = 707 mK, Condenser = 155 mbar, IVC = 25 nbar Close 1K pot needle valve and 1K pot speedivalve. Set MC/Still heaters = 20/20 mW. |
| 1410 | Tower = 3.4 K, Condenser = 328 mbar, IVC = 25 nbar |
| 1450 | Tower/1K Pot/Sorb = 6.57/4.32/5.5 K, Condenser = 328 mbar, IVC = 34 nbar. Condenser rising, Tower rising slowly |
| 1515 | Tower/1K Pot/Sorb = 7.21/6.35/8.1 K, Condenser = 560 mbar, IVC = 46 nbar. Condenser and Tower look flattish now. This is warmer than we got the Tower to last time, so I guess that means less helium in the IVC but it's still apparently present because the Tower temperature is hanging. And the IVC pressure is not as high as last time. |
| 1555 | Tower/1K Pot/Sorb = 7.26/6.54/8.4 K, Condenser = 580 mbar, IVC = 46 nbar. All temperatures and Condenser pressure flat. |
| let's try turning off heaters and see if this makes Pot and Sorb get warmer, like it did last time | |
| 1600 | Turn off MC and Still heaters. Tower temperature goes down - also Pot and Sorb go down. (Main Bath higher this time, 62%, maybe that's the difference). IVC falls to 34 nbar |
| 1650 | Tower/1K Pot/Sorb = 5.64/6.55/8.5 K, Condenser = 575 mbar, IVC = 34 nbar. Tower is not settling toward 4.4 K as quickly as on Wednesday |
| 1650 | Turn on MC and Still heaters again to 20/20 mW Tower bounces right back to 7.23 K, pot/sorb don't change much. IVC = 46 nbar |
| 1745 | Open Turbo valve to pump on IVC. Tower temperature starts increasing immediately. |
| 1750 | Fill LN2 cold trap bath |
| 1755 | Tower = 7.6 K |
| 1835 | Tower = 9.9 K |
| 1900 | Tower = 11.7 K, Sorb 11 K, 1K Pot > 7K |
| 1905 | Reduce heaters on MC/Still 20/20→3/3 mW |
| 1940 | Tower = 11.8 K and flat or falling slowly |
| 2140 | Tower = 11.75 K |