UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20180723

2018/07/23 (Monday)

Previous Day

overnight: Tower temp, circulation pressures, 1K Pot mostly flat
0805 Tower = 94 K, G1/G2 = 457/417 mbar, Still = 0.055 mbar / 0.1 mW, 1K Pot = 1.45 K / 4.0 mbar.
0810 Open 1K Pot needle valve 140→150°, to see what happens. Tower falls a little, Condenser rises a little.
0815 Turn off MIDAS DAQ programs, ground HV supply, power down all three DCRCs. (this causes a small spike in temperature)
0815 1K Pot = 1.61 K / 20 mbar
0820 Tower = 86 mK, Condenser = 483 mbar. No improvement. We should warm up again and see if there's as much pressure in the IVC as before.
0820 Turbo on IVC = 2.5 nbar. Close valve to turbo pump. With pump valved out, IVC = 25 nbar.
0923 Fill LN2 cold trap full.
filling trap caused fall in Condenser 480-455 mbar, rise in Tower temperature 89→103 mK
1050 Turn Still heater on to 5 mW, Still→0.110 mbar. Tower falls to 90 mK, at least initially. Condenser pressure rises above 500 mbar.
1235 Tower temperature started increasing again, now Tower = 120 mK, Condenser = 711 mbar.
1240 Suspend circulation: close V6, open V9
1310 Tower = 707 mK, Condenser = 155 mbar, IVC = 25 nbar
Close 1K pot needle valve and 1K pot speedivalve. Set MC/Still heaters = 20/20 mW.
1410 Tower = 3.4 K, Condenser = 328 mbar, IVC = 25 nbar
1450 Tower/1K Pot/Sorb = 6.57/4.32/5.5 K, Condenser = 328 mbar, IVC = 34 nbar.
Condenser rising, Tower rising slowly
1515 Tower/1K Pot/Sorb = 7.21/6.35/8.1 K, Condenser = 560 mbar, IVC = 46 nbar.
Condenser and Tower look flattish now. This is warmer than we got the Tower to last time, so I guess that means less helium in the IVC but it's still apparently present because the Tower temperature is hanging. And the IVC pressure is not as high as last time.
1555 Tower/1K Pot/Sorb = 7.26/6.54/8.4 K, Condenser = 580 mbar, IVC = 46 nbar.
All temperatures and Condenser pressure flat.
let's try turning off heaters and see if this makes Pot and Sorb get warmer, like it did last time
1600 Turn off MC and Still heaters. Tower temperature goes down - also Pot and Sorb go down. (Main Bath higher this time, 62%, maybe that's the difference). IVC falls to 34 nbar
1650 Tower/1K Pot/Sorb = 5.64/6.55/8.5 K, Condenser = 575 mbar, IVC = 34 nbar.
Tower is not settling toward 4.4 K as quickly as on Wednesday
1650 Turn on MC and Still heaters again to 20/20 mW
Tower bounces right back to 7.23 K, pot/sorb don't change much. IVC = 46 nbar
1745 Open Turbo valve to pump on IVC. Tower temperature starts increasing immediately.
1750 Fill LN2 cold trap bath
1755 Tower = 7.6 K
1835 Tower = 9.9 K
1900 Tower = 11.7 K, Sorb 11 K, 1K Pot > 7K
1905 Reduce heaters on MC/Still 20/20→3/3 mW
1940 Tower = 11.8 K and flat or falling slowly
2140 Tower = 11.75 K

Next Day