2018/07/22 (Sunday)

Previous Day

overnight: Tower slowly climbing, then around 11am at 37 mK it climbed faster and steadily. Condenser pressure also rose steadily, about 100 mbar over 8 hours. Still and 1K Pot both pretty steady
1441 Tower = 57.5 K, G1/G2 = 434/395 mbar, P1 = 0.055 mbar, 1K Pot = 1.467 K / 4.5 mbar.
1441 Fill LN2 cold trap full.
1454 Open 1K Pot needle valve about 10 degrees. Tower temp peaks at 67.0 mK, 1K pot pressure goes offscale.
1459 Turn on still heater to 0.1 mW
1500 Start LHe transfer (MB = 28.2%)
1535 1K Pot = 1.55 K / 13 mbar
1606 End LHe transfer (MB = 89.8%)
1607 1K Pot = 1.47 K / 5.5 mbar
1605 Tower = 78 mK and rising, not good
1610 Try turning on Still heater to 3 mW. Tower falls a bit, but Condenser pressure rises more quickly than before.
1621 Tower = 75.7 mK, G1/G2 = 484/443 mbar, P1 = 0.095 mbar, 1K Pot = 1.466 K / 5.0 mbar.
1715 Tower has flattened out around 80 mK, Condenser flattening at 550 mbar
2035 Tower temperature is climbing again, condenser pressure rising slowly too.
Tower = 86 mK, Condenser = 580 mbar
2040 Try raising Still heater 3→6 mW. Still 0.090→0.130 mbar.
Tower temp dipped, then bounced back, then started slowly falling a bit.
2110 Tower = 85 mK
2140 Tower temperature climbing again, now 92 mK. Condenser has risen to 690 mbar.
This is not good.
2205 Lower Still heater 6.0→0.1 mW. Pressure comes down as I hoped. Tower temperature rises, not surprising. Over 100 mK now.