2018/07/20 (Friday)

Previous Day

Status: circulation and 1K Pot suspended. Mixture in dump. MC and Still heaters on at 3 mW each. Pumping out IVC with turbo pump overnight.
overnight: Tower steady at 13 K, Sorb at 16K
0740 Turn off MC and Still heaters
0745 Turbo on IVC = 2.5 nbar
0810 Open V9 and V13a to let mix in dump cool in Trap 1. Open V12a + speedivalve
0815 Fill LN2 cold trap bath
0815 Open V1 and V3 to let mixture into fridge. Condenser speedivalve and Still gate valve are open. 3He pump is off.
G1/G2 → 540/530 mbar
1110 Tower/Sorb/1K Pot = 11/9.4/>7 K. Sorb is now flat, Tower falling steadily. G1/G2 = 570/525 mbar, flat.
1115 IVC = 25 nbar with Turbo valved out. Open valve to Turbo again, we'll pump IVC continuously.
1120 Start pumping on 1K Pot. Alternate between opening 1K Pot speedivalve and V4a, to not overstress pump. Pressure ~ 0.2 bar first time speedivalve opened.
With both valves open, 1K Pot/Sorb → 4.8/6.2 K
1125 1K Pot = 4.3 K / 0.65 mbar
Open 1K Pot needle valve → 10°. 1K Pot → 3.4 K / 3.4 mbar
Open 1K Pot needle valve → 20°. 1K Pot → 2.9 K / 6.0 mbar
1130 Open 1K Pot needle valve → 30°. 1K Pot → 2.7 K / 9.0 mbar
1135 Open 1K Pot needle valve → 40°. 1K Pot → 2.4 K / 10.0 mbar
Open 1K Pot needle valve → 50°. 1K Pot → 2.4 K / 14.0 mbar
1140 Open 1K Pot needle valve → 60°. 1K Pot → 2.2 K / 17.0 mbar
1140 Tower = 10.1 K, Condenser = 472 mbar, both falling
1150 Tower = 3.4 K (fell fast). Condenser = 435 mbar
1K Pot = 2.7 K / 8 mbar. Pressure has been falling.
1150 Open 1K Pot needle valve → 70°. 1K Pot → 2.4 K / 17.0 mbar
1155 Open 1K Pot needle valve → 80°. 1K Pot → 2.2 K / 26.0 mbar
1200 Tower = 3.1 K, Condenser = 362 mbar, both falling
1220 Fill LN2 cold trap full.
1230 Tower = 2.5 K, Condenser = 142 mbar.
1K Pot = 1.9 K / 5.5 mbar.
Open 1K Pot needle valve → 90°. 1K Pot → 1.6 K / 18.0 mbar
1250 Tower = 2.1 K, Condenser = 61 mbar.
1K Pot = 1.5 K / 5.0 mbar.
Open 1K Pot needle valve → 100°. 1K Pot → 1.6 K / 9.0 mbar
1303 Start LHe transfer (MB = 37.5%)
1315 Tower = 1.88 K, Condenser = 49 mbar.
1K Pot = 1.4 K / 4.5 mbar.
Open 1K Pot needle valve → 110°. 1K Pot → 1.5 K / 7.5 mbar
1330 Tower temperature suddenly dips from 1.85→1.75 K. Condenser pressure flattens out a bit.
1335 Tower = 1.73 K, Condenser = 38 mbar.
1K Pot = 1.4 K / 4.5 mbar.
Open 1K Pot needle valve → 120°. 1K Pot → 1.6 K / 22 mbar
1355 Tower = 1.70 K and flattish, Condenser = 32 mbar and still falling. 1K Pot = 1.5 K / 6.5 mbar.
1355 End LHe transfer (MB = 89.9%)
1415 Tower = 1.67 K, Condenser = 29 mbar. 1K Pot = 1.5 K / 5.0 mbar.
1420 Turn on 3He pump. This pushes mixture that's behind pump into the fridge. G2 rises a little.
1430 Tower = 1.67 K, Condenser = 28 mbar. 1K Pot = 1.5 K / 5.0 mbar.
Open 1K Pot needle valve → 130°. 1K Pot → 1.6 K / 16 mbar
1430 Close V9, leaving 33 mbar of mixture in dump. Close V3.
1435 Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 75 mbar. Tower falls to 1.6 K.
1440 V6 → 13.2%. Condenser peaks at 108 mbar. Tower → 1.51 K
1450 Tower temperature is actually rising slowly, 1.5→1.52 K
1440 V6 → 13.6%. Condenser peaks at 220 mbar. Still → 40 mbar. Tower → 1.41 K
1505 V6 → 14.0%. Condenser peaks at 280 mbar. Still → 3.4 mbar. Tower → 1.06 K
1515 V6 → 14.4%. Condenser continues to fall. Still → 1.8 mbar. Tower → 0.82 K
1520 V6 → 15.0%. Condenser peaks at 166 mbar. Still → 0.75 mbar. Tower → 0.34 K
1535 V6 → 100%
1555 Tower = 42.4 mK, Condenser = 122 mbar and falling, Still = 0.065 mbar, 1K Pot = 1.47 K / 5.5 mbar
1625 Tower = 31.0 mK, Condenser = 106 mbar, Still = 0.060 mbar, 1K Pot = 1.48 K / 5.5 mbar
1625 Open V14 to pull the rest of the mixture into circulation. Still peaks at 16 mbar then falls.
Condenser peaked at 145 mbar then fell, Tower peaked at 51 mK then fell, then bounced up a bit to 33 mK
1650 Fill LN2 cold trap bath
1650 Still = 0.7 mbar, low enough: close V14
1710 Tower = 29.4 mK, Condenser = 113 mbar, Still = 0.070 mbar, 1K Pot = 1.48 K / 5.0 mbar
1730 Tower = 29.0 mK, Condenser = 123 mbar, Still = 0.070 mbar, 1K Pot = 1.48 K / 5.0 mbar
1930 Tower = 29.6 mK, Condenser = 118 mbar, Still = 0.060 mbar, 1K Pot = 1.47 K / 4.5 mbar