2018/07/13 (Friday)
| Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3) | |
| 1244 | Fill LN2 cold trap full. |
| 1245 | Tower = 33.2 mK, G1/G2 = 154.8/145.9 mbar, P1 = 0.140 mbar (Still heater = 4 mW), 1K pot = 1.632 K / 28 mbar. MB = 27.3%. |
| 1425 | Start LHe transfer (MB = 23.5%). |
| 1516 | End LHe transfer (MB = 87.0%). |
| 1526 | Tower = 33.2 mK, G1/G2 = 146.4/138.5 mbar, P1 = 0.140 mbar (Still heater = 4 mW), 1K pot = 1.614 K / 28 mbar. MB = 86.0%. |