2018/07/07 (Saturday)
| Status: DCRCs off. Trying to cool to base but getting stuck around 1K | |
| 0925 | Tower = 1.6 K. Condenser = 342 mbar. V6 → 14.0%. |
| 1145 | Tower = 1.35 K. Condenser = 532 mbar, plateauing. Still = 5.5 mbar. Turn on Still heater to 5 mW |
| 1230 | Tower = 1.45 K. Condenser = 693 mbar, peaking. Still = 9.5 mbar. Turn Still heater back down to 0.1 mW |
| 1430 | Tower = 1.35 K. Condenser = 516 mbar and falling. Still = 6.0 mbar. |
| 1615 | Fill LN2 cold trap full. |
| 1615 | Tower = 1.35 K, G1/G2 = 432.6/391.3 mbar, 1K Pot = 1.692 K / 70 mbar. |
| 1630 | Tower = 1.35 K. Condenser = 407 mbar. Still = 9.5 mbar. Main bath = 14.2% |
| 2050 | For the last few hours, Condenser pressure is oscillating with a 20 minute period. Tower temperature also oscillating between 1.2 and 1.4 K. Looks like they are oscillating in phase. Still pressure oscillating between 10 and 40 mbar. |
| 2055 | V6 → 14.4% |
| 2115 | Tower = 1.1 K. Condenser = 235 mbar, settling. Still = 4.0 mbar. Oscillations seem to have stopped. |
| 2140 | Tower = 1.2 K. Condenser = 242 mbar. Still = 4 mbar. V6 → 15%. Condenser peaks at 265 mbar |
| 2205 | Tower = 405 mK and falling, Condenser = 239 mbar and falling, Still = 1.6 mbar. |
| 2235 | Tower = 286 mK and flat, Condenser = 210 mbar and falling, Still = 1.3 mbar. |
| 2255 | Tower = 274 mK, Condenser = 203 mbar, Still = 1.2 mbar. V6 → 100%. Condenser peaks at 230 mbar |
| 2310 | Tower = 94 mK, Condenser = 209 mbar, both falling. Still = 0.15 mbar. |
| 2315 | Tower = 46 mK, Condenser = 177 mbar, both falling. Still = 0.13 mbar. |
| well this is annoying - now it's working, but we decided to skip the transfer because I thought we'd need to warm up | |
| 2315 | MB = 2.7% |
| 2330 | Tower = 42 mK, now kind of flat. Condenser = 162 mbar and falling. Still = 0.12 mbar. Increase Still heat to 2 mW |
| 2340 | Tower seems stuck exactly at 42.24 mK. I think that means the resistance is higher than the range on the bridge, which I left at 67 kΩ, should increase it to 200 kΩ |
| 0014 | Tower = 28.7 K, G1/G2 = 175.8/362.8 mbar, 1K Pot = 1.660 K / 38 mbar. |