2018/07/05 (Thursday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
2400 Tower = 35.9 mK, G1/G2 = 321/290 mbar , Still = 0.050 mbar / 1 mW. 1K Pot = 1.38 K / 2.0 mbar. It won't last till morning.
0047 Open 1K Pot needle valve about 10 degrees. 1K Pot peaks at 2.15 K and falls / >1000 mbar.
overnight: 1K Pot pressure never fell below “1000 mbar” (except for a period between 3:30 and 4:30 am, but it was still high). 1K Pot temperature remained above 1.7 K. Condenser pressure remained in control so things are basically working, but Tower temperature is elevated.
Should close the needle valve a little when we get in this morning.
0730 Tower = 51.3 mK, G1/G2 = 473/424 mbar , Still = 0.080 mbar / 1 mW. 1K Pot = 1.76 K / “1000” mbar.
0740 Reduce Still heat from 1 mW to 0.1 mW. Tower temperature and circulation pressures fall. Tower settles at 45.9 mK
0935 Close 1K Pot needle valve a few degrees. 1K Pot temperature goes up instead of down, 1.76→1.9 K, and pressure still pegged. But Tower temperature starts to fall.
1K Pot pressure is low on mechanical gauge. But 4He pump is straining a bit, vapor coming out of it.
0940 Close 1K Pot needle valve a few degrees more. Little change.
0945 Close 1K Pot needle valve about 10 degrees more. Pot temperature goes down a little, → 1.87 K
0950 Close 1K Pot needle valve about 10 degrees more. Pot temperature → 1.83 K
0955 Close 1K Pot needle valve about 10 degrees more. Pot temperature → 1.78 K and falls slowly.
1005 Close 1K Pot needle valve about 10 degrees more. Pot temperature spikes up to 1.83 K but then falls to 1.78 K
1010 Close 1K Pot needle valve about 10 degrees more. Pot temperature → 1.73 K, pressure falls into readable range on P2
1020 Tower = 44.8 mK. Turn Still heater up to 1 mW again.
1025 1K Pot = 1.73 K / 260 mbar. Pressure falling slowly.
1030 1K Pot = 1.73 K / 300 mbar, kind of hanging there
Close 1K Pot needle valve about 10 degrees more. 1K Pot → 1.71 K / 110 mbar
1035 Condenser pressure = 456 mbar, starts falling
1035 MB = 32.3%
1040 Fill LN2 cold trap bath
1K Pot is still pretty warm, condenser pressure still high-ish (450 mbar), Tower temperature still high-ish (47 mK). Might want to stop circulation and start over.
1040 1K Pot = 1.70 K / 80 mbar, kind of hanging there
Close 1K Pot needle valve about 10 degrees more. 1K Pot → 1.66 K / 38 mbar. Condenser = 429 mbar, falling.
1045 1K Pot = 1.63 K / 26 mbar. G1/G2 = 423/376 mbar, now flat. Tower = 50.5 mK and actually increasing, that's bad. Still = 0.065 mbar.
1050 Increase Still heater 1→2 mW. Still → 0.075 mbar, Tower 52→51.4 mK.
1210 Reduce Still heat 2→1 mW
1210 Tower = 61.4 mK (yuk), G1/G2 = 436/388 mbar , Still = 0.060 mbar / 1 mW. 1K Pot = 1.51 K / 6.0 mbar.
Pot behaving pretty normally now but base sucks, condenser pressure high.
1210 Close V6 to stop circulation. Still pressure rises, Condenser falls, Tower rises.
1220 Close 1K Pot needle valve, then close 1K Pot speedivalve to stop pumping on pot.
1220 Just noticed that we have not been pumping on the IVC since we warmed above 4K last week.
IVC = 34 nbar
1230 Still off-scale, Condenser = 265 mbar and falling, Tower = 0.74 K, 1K Pot = 2.6 K. IVC = 34 nbar.
1305 Condenser = 152 mbar and falling, Tower = 1.4 K, 1K Pot = 3.9 K. IVC = 34 nbar
1305 Open V3 to let Condenser side open to Still side. Condenser pressure falls, indicating that equilibrium had not yet been reached - might suggest poor throughput. Condenser → 88 mbar.
1310 Turn on MC heater to 20 mW. Tower temperature rises.
1325 Condenser = 214 mbar and rising, Tower = 2.9 K. 1K Pot = 4.1 K, flat. IVC still flat at 34 nbar.
1340 Condenser = 353 mbar and rising, Tower = 3.2 K. IVC = 34 nbar.
1345 Turn Still heater up to 20 mW
1353 Start LHe transfer @ MB = 28.2%
1400 Condenser = 655 mbar and rising, Tower = 3.8 K. IVC = 25 nbar. IVC went down, surprisingly - maybe because we started the transfer?
Starting the transfer seems to have stopped the rise of the Tower temperature and the Condenser pressure.
1420 Condenser = 680 mbar and flattish, Tower = 3.78 K and flattish. IVC = 15 nbar, even lower.
1445 Tower = 3.9 K, then suddenly starts climbing, to 5.3 K in 5 minutes. IVC → 63 nbar.
1450 Stop LHe transfer @ MB = 88.8%
1455 Condenser = 859 mbar and flattish, Tower = 5.46 K and flattish. IVC = 74 nbar. Start pumping out IVC with turbo pump.
1500 Tower = 5.6 K, Condenser = 893 mbar. Turn off MC and Still heaters
1520 Tower = 4.46 K and settling, Condenser = 903 mbar. Turbo on IVC = 2.5 nbar
1520 Open 1K Pot speedivalve to start pumping on pot. Pot dips to 1.9 K then rises back to 3.7 K. Pressure falls to 0.4 mbar. Condenser pressure falls to 743 mbar, Tower temperature falls to 3.8 K.
1525 Open 1K Pot needle valve → 10 degrees. 1K Pot → 3.35 K / 4.5 mbar
Open 1K Pot needle valve → 20 degrees. 1K Pot → 2.98 K / 6.0 mbar
1535 Open 1K Pot needle valve → 30 degrees. 1K Pot → 2.76 K / 9.5 mbar
Open 1K Pot needle valve → 40 degrees. 1K Pot → 2.28 K / 19 mbar
1540 Condenser = 431 mbar, Tower = 3.39 K, both falling
1540 1K Pot = 1.70 K / 19 mbar (pressure steady but temperature falling)
1550 Condenser = 131 mbar, Tower = 2.60 K, both falling
1550 1K Pot = 1.54 K / 11 mbar
1550 Open 1K Pot needle valve → 50 degrees. 1K Pot → 1.54 K / 11 mbar (not much change)
Open 1K Pot needle valve → 60 degrees. 1K Pot → 1.55 K / 12 mbar (not much change)
1555 Condenser = 84 mbar, Tower = 2.38 K, levelling out
1555 Close V3
1555 Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 170 mbar. Tower falls to 2.23 K. 1K Pot kinda flat…
1600 Open 1K Pot needle valve → 70 degrees. Not much change.
1605 V6 → 13.2%. Condenser rises above 200 mbar. 1K Pot spike → 1.71 K / off-scale, settles at 1.68 K / 75 mbar. Tower → 2.14 K.
1615 Condenser finally started falling after reaching max 210 mbar. Tower = 1.99 K
1620 V6 → 13.5%. Condenser eventually peaks at 350 mbar. Tower → 1.68 K. 1K Pot = 1.38 K / 75 mbar and steady.
1640 V6 → 13.8%. Condenser peaks at 405 mbar. Tower → 1.41 K. Still → 8 mbar.
1650 V6 → 14.2%. Condenser peaks at 420 mbar. Tower → 1.29 K. Still → 3.2 mbar.
1655 Fill LN2 cold trap bath
1655 V6 → 15.0%. Condenser peaks at 420 mbar. Tower → 1.30 K (wavering). Still → 0.55 mbar.
1710 V6 → 100%. Condenser rises to 405 mbar quickly, then continues to rise more slowly but steadily. Tower → 1.31 K, flattish. Still → 0.095 mbar.
not sure what to make of this. The condenser has risen to 417 mbar and keeps climbing linearly. Tower seems stuck around 1.3 K. Still is pretty low.
this is a little different from the usual cooldown because no mixture was left in the dump, we started with all of it.
1720 Tower finally starts cooling again…
1725 Tower = 1.15 K and falling. Condenser = 423 mbar and no longer rising.
1730 Tower = 1.07 K and falling. Condenser = 422 mbar
1740 Still = 0.080 mbar. Turn on Still heater → 1.0 mW
1900 Tower = 0.99 K and seems to be oscillating with a ~3 min period. Condenser = 432 mbar. Still = 0.085 mbar. 1K Pot = 1.68 K / 55 mbar.
1900 Increase Still heater 1 → 5 mW
1955 Tower rose to 1.09 K, now down to 1.04 K. Condenser has risen to 484 mbar and continues to rise. Still has only risen a little, 0.110 mbar. Doesn't look good.
1955 Decrease Still heater → 0.1 mW
2348 Tower = 0.92 K and oscillating, Condenser = 423 mbar, 1K Pot = 1.67 K / 45 mbar. Try closing to V6=13.4%