2018/07/04 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
overnight: base temperature has been wavering between 32 and 35 mK, not as stable as before. Seems to have changed since yesterday's transfer. Our last problem with base temperature seemed to start after a transfer too. In that case the stick was inserted almost completely after almost no pre-cool. During yesterday's transfer the stick was lowered somewhat early during the precool rather than waiting with it just barely inserted. It could be that these thermal shocks are producing problems.
0515 Tower = 34.8 mK, G1/G2 = 287/262 mbar , Still = 0.055 mbar / 1 mW. 1K Pot = 1.44 K / 4.0 mbar
1319 Tower = 35.9 mK, G1/G2 = 289/263 mbar , Still = 0.050 mbar / 1 mW. 1K Pot = 1.38 K / 2.2 mbar
1320 Fill LN2 cold trap full.
1325 Open 1K Pot needle valve about 10 degrees. Tower temp peaks at 42.5 mK and begins to fall, 1K pot pressure goes offscale.
1338 Tower temp falling (33.1 mK), 1K Pot = 1.835 K and falling / >1000 mbar.
2400 Tower = 35.9 mK, G1/G2 = 321/290 mbar , Still = 0.050 mbar / 1 mW. 1K Pot = 1.38 K / 2.0 mbar. It won't last till morning.