2018/07/03 (Tuesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
overnight: 1K pot drifting down. At 06:45 it failed: 1K Pot fell to 1.38 K / 1.45 mbar, then Pot temperature started rising, then at 6:50 Pot temperature and circulation/still pressures began rising.
0740 G1/G2 = 800/750 mbar, 1K Pot = 4 K. Close V6 and open V14 to stop circulation and relieve pressure. G1/G2 begin to fall.
0900 Tower = 1.57 K, Still pressure offscale / 1.5 mW, G1/G2 = 108/96 mbar, V6/V14 = closed/open, 1K Pot = 2.14 K / 1.4 mbar. MB = 35.2%
0900 Open 1K Pot needle valve about 20 degrees more. 1K Pot → 1.7 K / 70 mbar then drifts down. Condenser pressure falls.
0905 Detector bias was left at 50 V. Run script changeHV-50_0.msl to dial it down to 0 V.
0905 Turn off Still heat from 1.5 mW to 0 mW.
0910 1K Pot pressure falls to 10 mbar then spontaneously spikes and falls again - pot goes into oscillatory mode.
0930 1K Pot = 1.62 K / 22 mbar. G1/G2 = 60/52 mbar and flat. Tower = 1.63 K
0930 Close V14 leaving 53 mbar mixture in dump
0930 Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 110 mbar.
0945 V6 → 13.2%. Condenser peaks at 67 mbar. Tower → 1.53 K.
0950 V6 → 13.5%. Condenser peaks at 160 mbar. Tower → 1.43 K. Still falls into readable range.
1000 Fill LN2 cold trap full.
1005 V6 → 13.8%. Condenser peaks at 230 mbar. Tower → 1.28 K. Still → 12 mbar.
1015 V6 → 14.2%. Condenser peaks at 200 mbar. Tower → 0.97 K. Still → 2.4 mbar.
1020 V6 → 15.0%. Condenser peaks at 180 mbar. Tower → 0.86 K. Still → 0.75 mbar.
1030 V6 → 100%. Condenser peaks at 185 mbar. Tower → 0.42 K. Still → 0.12 mbar.
1125 Tower = 34.4 mK, G1/G2 = 118/111 mbar, Still = 0.080 mbar, all pretty flat. 1K Pot = 1.51 K / 6 mbar
1125 Open V14 to let rest of mix into circulation. Tower peaks at 115 mK, Condenser peaks at 210 mbar, then both begin to fall.
1200 Tower = 38.4 mK and falling, G1/G2 = 113/107 mbar and flat, Still = 0.5 mbar
1200 Close V14
1205 Tower = 34.8 mK and falling, G1/G2 = 113/106 mbar and flat, Still = 0.080 mbar. 1K Pot = 1.51 K / 5.5 mbar
1205 Turn on Still heater to 1.0 mW
1230 Tower = 30.5 mK and settling, G1/G2 = 127/121 mbar , Still = 0.100 mbar / 1 mW. 1K Pot = 1.50 K / 5.5 mbar
1455 Tower = 30.6 mK, G1/G2 = 123/117 mbar , Still = 0.095 mbar / 1 mW. 1K Pot = 1.45 K / 4.0 mbar
1502 Start LHe transfer @ MB = 26.4%
1544 End LHe transfer @ MB = 91.7%
Transfer caused a brief spike in Tower temperature 30.6 → 32.3 mK then a significant drop → 26.8 mK but then it came back up as the transfer ended.
Condenser pressure rose significantly during transfer.
1615 Tower = 32.1 mK, G1/G2 = 232/216 mbar , Still = 0.080 mbar / 1 mW. 1K Pot = 1.41 K / 3.8 mbar
1650 Fill LN2 cold trap bath
1655 Tower = 32.2 mK, G1/G2 = 227/209 mbar , Still = 0.080 mbar / 1 mW. 1K Pot = 1.41 K / 3.6 mbar
1655 Open 1K Pot needle valve another ~15 degrees further. Temperature spikes to 1.9 K, pressure goes off scale.
1700 Tower = 30.9 mK, G1/G2 = 262/242 mbar , Still = 0.110 mbar / 1 mW. 1K Pot = 1.72 K / 400 mbar, falling
1705 Tower has increased back to 32 mK. Since the base and condenser pressure are somewhat higher since the transfer I wonder if the temperature jolt from the transfer changed things (maybe knocking some snow loose and causing a partial blockage).
1710 Tower = 32.0 mK, G1/G2 = 279/257 mbar , Still = 0.110 mbar / 1 mW. 1K Pot = 1.67 K / 50 mbar
2322 Tower = 32.9 mK, G1/G2 = 251/230 mbar , Still = 0.050 mbar / 1 mW. 1K Pot = 1.40 K / 2.8 mbar
0032 Tower = 33.1 mK, G1/G2 = 253/231 mbar , Still = 0.050 mbar / 1 mW. 1K Pot = 1.39 K / 2.2 mbar. Open 1K Pot needle vlave +20 degrees. 1K Pot = 1.8 K and falling / >1000 mbar
0042 1K Pot = 1.73 K / 450 mbar and falling