2018/06/12 (Tuesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
0345 Tower = 96 mK, Still = 0.100 mbar, G1/G2 = 421/390 mbar, 1K Pot = 1.85±0.10 K / 1.7 mbar.
Tower temperature was around 100 mK for several hours but now starting to fall… This coincided with a small increase in the 1K pot pressure from 0.9 mbar to 2 mbar and sudden increase in 1K pot pressure from stable 1.36 K to relatively unstable 1.8±0.10 K
0415 Tower = 84 mK, lowest it has been in last 12+ hours, and falling. Good sign I guess. (Maybe we had a partial blockage and it came loose. Sure, why not?)
0615 Tower = 67 mK, still falling
0820 Tower = 61 mK, still falling
0838 Tower/MC/Coldplate =60.3/58.9/190 mK, Still = 1.26 K / 0.10 mbar, G1/G2 = 353/329 mbar, 1K Pot = 1.6±0.1 (somewhat unstable) K / 2.2 mbar. MB = 73.8%
0844 Open 1K pot +10 degrees. P2 → 8 mbar and starts to decay, 1K pot temperature stabilizes at 1.55 K and slowly decreases
0906 1K pot spontaneously jumps and oscillates around 1.57 K / 12 mbar. Tower = 59.6 mK
0924 Tower = 60.3 mK, G1/G2=348/323 mbar, 1K pot = 1.54 K / 8.5 mbar. Still heater = 3.5 mW, P1 = 0.10 mbar
0925 Let's play with the Still heating and see if that helps things. Turn Still heater off.
0930 P1 falls to 0.045 mbar, Tower rises to 63.1 mK.
0931 Turn Still heater back on to 3.5 mW
0936 P1 → 0.045 mbar, Tower → 55.3 mK and decreasing, G1/G2 → 322/300 mbar
0938 1K pot is slowly falling as usual to 1.52 K / 7 mbar
0951 Tower bottomed out at 48.2 mK, now P1 → 0.10 mbar, Tower → 50.2 mK G1/G2 → 332/308 mbar, 1K pot = 1.49 K / 5.0 mbar. Increase Still heating to 5.0 mW
0955 Tower temp starts to increase to 53 mK, P1 rises to 0.12 mbar. Turn Still heater back to 3.5 mW
0956 Fill LN2 cold trap.
1000 All Linux machines freeze. :/
1010 Anthony suggested stopping/starting circulation could loosen the suspected partial fridge blockage. Let's try that. G1/G2 = 321/295 mbar. Close V13A to stop circulation gently.
1015 G1/G2 = 215/406 mbar, reopen V13A. G1/G2 → 334/313 mbar. That didn't seem to do anything. But we didn't really change the pressure at the impedance that much.
1026 Tower = 65.3 mK
1030 Let's try a more drastic circulation restart by pumping the mix to the dump from both sides of the fridge like we're going to warm up, then restarting circulation from scratch. Close V13A, open V9, open V5. G1/G2 → 6.6/80 mbar, Tower → 120 mK
1034 Close V5, V9. Close V6 fully. G1/G2 = 12.9/120 mbar. G2 rising presumably as cold mix comes to room temp.
1040 Watch V6 stepper motor as it is closing. Notice that the motor shifts slightly each step because the screws that attach the motor to the valve body are loose. This may be the source of some of the V6 valve slop. Tighten those screws with an allen key.
1042 Restart circulation by opening V13A, V14 and opening V6 → 12.4%. G1/G2 →95/96 mbar, P1 → >1000 mbar, Tower → 0.99 K
1047 V6 → 13%. G1/G2 →95/96 mbar, P1 → >1000 mbar, Tower → 0.99 K
1052 V6 → 13.5%. G1/G2 → 284/265 mbar, P1 → 32 mbar, Tower → 1.59 K
1110 G1/G2 flat at 285/265 mbar
1115 V6 → 14.0%. G1/G2 → 473/439 mbar, P1 → 5 mbar, Tower → 1.21 K. 1K pot oscillates a little around 1.4K / 3 mbar
1132 Close needle valve and immediately open back up to same position +10 degrees. P2 → >1000 mbar then eventually settles to 6.5 mbar with no oscillations
1158 G1/G2 = 538/497 and flattening out. P1 = 4.0 mbar, Tower = 1.24 K. I don't think we're getting any circulation now. Dang, we might have fully blocked it with the circulation restart…
1230 G1/G2 = 439/406 and slowly falling. P1 = 2.8 mbar. Tower = 1.25±0.2 K
1233 Still heater may not be helping things at this point. Turn off Still heater. P1 → 1.1 mbar, Tower → 1.05±0.2 K
1238 V6 → 14.4%. G1/G2 → 356/330 mbar and decreasing, P1 → 0.50 mbar, Tower → 1.0±0.1 K
1244 V6 → 14.8%. G1/G2 → 310/288 mbar and decreasing, P1 → 0.40 mbar, Tower → 0.90±0.1 K
1503 G1/G2 → 242/256 mbar and decreasing, P1 → 0.40 mbar, Tower → 0.90±0.1 K
1305 V6 → 15.2%. G1/G2 → 201/187 mbar and decreasing, P1 → 0.26 mbar, Tower → 0.57±0.05 K
1309 Close 1K pot needle valve and immediately reopen to same position + 10 degrees.
1319 1K pot = 1.58 K / 13 mbar
1320 V6 → 100%. G1/G2 → 127/121 mbar and decreasing, P1 → 0.090 mbar, Tower → 249 mK
1430 G1/G2 → 106/101 mbar and decreasing, P1 → 0.036 mbar, Tower → 155 mK. Close V14.
1435 1K pot = 1.38 K / 1.9 mbar. Open 1K pot needle valve +10 degrees.
1500 G1/G2 → 100/97 mbar and decreasing, P1 → 0.034 mbar, Tower → 121 mK, 1K pot = 1.44 K / 3.8 mbar. Seems to be cooling okay and it's encouraging that the fridge pressures are far down from what they were this morning.
1503 Let's try helping it out with Still heating. Turn on Still heater to 3.5 mW. P2 → 0.11 mbar, G1/G2 → rising, Tower → cooling faster
1554 G1/G2 → 303/283 mbar and starting to flatten out slowly, P1 → 0.10 mbar, Tower → 36.2 mK, 1K pot = 1.40 K / 2.5 mbar. Alright, we're back to base, but we still have a 1K pot that keeps running dry.
1556 Open 1K pot needle valve +10 degrees.
1614 G1/G2 → 326/304 mbar, P1 → 0.10 mbar, Tower → 43.9 mK, 1K pot = 1.49 K / 5.0 mbar. Tower has risen slightly.
1650 Maybe air is leaking down the 1K pot needle valve shaft when we turn it. Enclose needle valve in ziploc bag filled with He from the wall line. Open 1K pot needle vale +10degrees through bag.
1654 Fill LN2 cold trap full.
1700 Tower → 55.0 mK
1827 G1/G2 = 333/312 mbar, P1 = 0.10 mbar, Tower → 59.5 mK, 1K pot = 1.38 K / 2.4 mbar. Ziploc trick didn't seem to help.
2134 G1/G2 = 336/313 mbar, P1 = 0.10 mbar, Tower → 55.1 mK, 1K pot = 1.38 K / 2.0 mbar.
2326 G1/G2 = 437/403 mbar, P1 = 0.11 mbar, Tower → 70.4 mK, 1K pot = 2.33 K / 1.4 mbar. Crap, looks like the 1K pot is failing.
2347 Open 1K pot needle valve +20 degrees. G1/G2 start to fall, 1K pot temp → 1.62 K / 22 mbar
2350 1K pot pressure is already falling below 20 mbar. Open 1K pot needle valve +10 degrees. P2 rises to 100 mbar, decays down to 40 mbar over couple minutes, then runs away to >1000 mbar. 1K pot temperature is oscillating around 1.8 K
0001 P2 finally falls into readable range. It's falling steeply and oscillating slightly.
0003 G1/G2 = 413/381 mbar, P1 = 0.10 mbar, Tower → 77.6 mK, 1K pot = 1.71 K / 85 mbar (and falling).
0017 P2 has fallen to 20 mbar and is not oscillating. This still doesn't seem stable and likely won't last the night. Close V6 to stop circulation.
0022 Open V9 and V3, turn off 3He pump.