start › cdms › k100 › run_diary › 20180613

2018/06/13 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
0001 P2 finally falls into readable range. It's falling steeply and oscillating slightly.
0003 G1/G2 = 413/381 mbar, P1 = 0.10 mbar, Tower → 77.6 mK, 1K pot = 1.71 K / 85 mbar (and falling).
0017 P2 has fallen to 20 mbar and is not oscillating. This still doesn't seem stable and likely won't last the night. Close V6 to stop circulation.
0022 Open V9 and V3, turn off 3He pump.
0100 Tower has risen to 1.2 K
0949 Tower/MC/Coldplate = 1.66/1.83/3.0 K, Still = 2.4 K/>1000 mbar/3.5 mW, G1/G2 = 39.5/34.4 mbar, 1K Pot = 1.36 K/1.4 mbar, MB = 40.2%
0954 Fill LN2 cold trap full.
1116 Close 1K pot needle valve and re-open 1 full turn (about +45degrees from previous position). 1K pot → 2.6 K and falling / >1000 mbar
1120 Close V9, leaving 40.5 mbar in dump. Turn on 3He pump, close V3, open V6 → 12.4% to start circulation through trap 1.
1130 4He pump suddenly gets very loud, and is smoking a little bit. 1K pot = 2.5 K / >1000 mbar. Close needle valve 1/4 turn, no change. Close needle valve completely. 1K pot → 1.4 K / 4 mbar , but pump is still noisy. 1K pot line is also frosty past the speedivalve.
1133 P2 = 4 mbar. Close V4A momentarily and P2 → <0.1 mbar, but pump is still noisy. 4He pump oil level is also at 0.
1137 Oh wait, it's the 3He pump that's noisy!
1140 Let's do this more slowly. Open needle valve 20 degrees. 1K pot → 1.41 K / 2.8 mbar
1140 Condenser peaked at 60 mbar, but has fallen to 39 mbar since I slowed the 1K pot. V6 → 12.8%
1142 The pumps seem happier now. The 3He pump is around normal volume and the 3He pump is not smoking. The 1K pot line has also thawed. we must have gotten LHe way up in the pot or something and the 4He pump couldn't take it. Maybe that's a good sign though that the needle valve can be opened far enough to get decent flow?
1145 Condenser pressure is settling at 37 mbar. 1K pot = 1.40 K / 2.6 mbar.
1146 V6 → 13.2%. Condenser peaks at 137 mbar, Tower = 1.68 K, 1K pot = 1.40 K / 3.0 mbar
1155 Open 1K pot needle valve +10 degrees. 1K pot → 1.44 K / 3.8 mbar. P2 changed smoothly from 3 to 3.8 when I did this. It didn't go off scale like it has been.
1159 Condenser pressure has risen to 152 mbar since tweaking the 1K pot.
1215 Condenser pressure = 186 and rising but flattening off. 1K Pot = 1.46 K / 4.5 mbar and oscillating a little bit.
1304 Condenser pressure peaked at 196 mbar and is finally coming down. Condenser = 179 mbar, 1K pot = 1.41 K / 2.8 mbar
1305 V6 → 13.4%.
1346 Condenser pressure has flattened out around 360 mbar and P1 = 85.0 mbar
1350 Open 1K pot needle valve +10 degrees. P2 rises to 20 mbar, then starts to oscillate around 7 mbar. 1K pot temp = 1.5 K. Still pressure (P1) climbs to 800 mbar
1400 Condenser pressure = 360 mbar. V6 → 13.6%. Condenser pressure → 420 mbar , P1 → 15 mbar, Tower = 1.6 K
1410 Condenser pressure = 430 mbar , P1 → 9.5 mbar, Tower = 1.4 K, 1K pot = 1.49 K / 5.0 mbar
1440 Condenser pressure = 426 mbar, P1 = 8.5 mbar. V6 → 13.8%.
1512 Condenser pressure peaked at 454 mbar, now 438 mbar. P1 = 4.0 mbar, Tower = 1.3 K. V6 → 14.0%.
1520 Commence LHe transfer (MB=31.6%)
1601 Condenser = 420 mbar. V6 → 14.2%. G1 slowly rises to 427 mbar
1605 1K pot = 1.40 K / 3.4 mbar. Open needle valve +10 degrees. P2 peaks at 35 mbar, then start to settle to ~ 5 mbar
1618 Cease LHe transfer (MB = 91.1%)
1627 Condenser pressure = 463 mbar, slowly climbing. Tower = 1.2 K
1700 Fill LN2 cold trap full.
1731 Condenser = 468 mbar, 1K pot = 1.49 K / 6.0 mbar. Tower = 1.2 K. The mix in G1 doesn't seem to want to condense even though the 1K pot is at 1.5 K. I'm going to keep waiting for the condenser pressure to come down before opening V6 any more.
2014 1K Pot = 1.37 K / 2.26 mbar. Open Needle Valve +10 degrees. 1K Pot → 1.62 K and falling / 25 mbar and falling
2026 Tower Temp = 1.05 K G1/G2 Stable at 472/432
2037 1K Pot still dropping, though it looks like it may be levelling off. 1K Pot → 1.57 K / 12.5 mbar
2157 Condenser pressure = 473 mbar and flat. 1K pot = 1.48 K / 5.5 mbar. Open Needle Valve +10 degrees. P2 rises to 45 mbar then starts falling.
2203 G1 = 475 mbar, P1 = 1.5 mbar, Tower = 1.1 K. The mix just isn't condensing. Maybe I'm trying to condense too much too fast? Let's slow circulation a bit. Reduce V6 to 13.4 %. G1 starts to fall slowly, P1 increases to 6 mbar
2209 Open 1K pot needle valve +10 degrees again. 1K pot pressure rises to 75 mbar before starting to oscillate around 30 mbar. 1K pot temp = 1.64 K
2213 G1 is flattening out to 430 mbar. The Still heater is at 3.5 mW. Let's try turning this off till we're colder. P1 falls from 8 mbar to 5 mbar
2215 1K pot= 1.60 K / 19 mbar.
2217 Leave it like this overnight. Condenser pressure is falling slowly, so hopefully we're actually condensing mix with this restricted circulation (V6=13.4%). Once the condenser pressure is really low in the morning, we can open V6 a little bit more.
2316 G1 = 340 mbar, P1 = 4.0 mbar, P2 = 6.0 mbar.