start › cdms › k100 › run_diary › 20180611

2018/06/11 (Monday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
Weekend: Circulation stopped. 1K pot still pumping. Tower temp around 1.6K
0929 Tower/MC/Coldplate =1.63/2.1/2.6 K, Still = 2.1 K / - mbar, G1/G2 = 28/30 mbar, 1K Pot = 1.38±0.01 K / 1.8 mbar. MB = 37.2%
0935 Fill LN2 cold trap.
0940 Wiggle needle valve back and forth +/- 10 degrees, but leave it in the same position it started in. P2 starts slow oscillation around 15 mbar.
0945 Close V9, leaving 30.2 mbar in dump. Turn on 3He pump, close V3, open V6 → 12.8% to start circulation through trap 1.
0950 Condenser pressure peaks at 94 mbar, Tower = 1.63 K
0955 1K pot = 1.50 K / 5.5 mbar
0957 V6 → 13.2%. Condenser peaks at 126 mbar. Tower → 1.45 K. Still pressure falls into readable range.
1023 V6 → 13.6%. Condenser rises 230 mbar. Tower → 1.32 K.
1037 1K pot = 1.44 K / 3.6 mbar
1040 Open 1K pot +10 degrees. 1K Pot → 1.6 K / 20 mbar.
1047 Condenser continues to rise slowly to 250 mbar.
1229 Condenser peaks at 270 mbar, falls to 220mbar. Tower → 1.32 K. V6 → 14.0%. Condenser rises 370 mbar. Tower → 1.12 K.
1233 P2 = 3mbar. Open 1K pot +10 degrees.
1250 V6 → 14.4%. Condenser keeps settling to 142 mbar. Tower → 900 mK.
1317 V6 → 14.8%. Condenser keeps settling to 130 mbar. Tower → 783 mK.
1330 Open V6 → 100%, Condenser settles to 128 mbar, Tower → 581 mK, Still = 0.06 mbar.
1345 Tower = 500 mK, 1 K pot = 1.49 K / 5.5 mbar
1350 Open V14 to let the rest of the mix into circulation. Condenser/Still peak at 180/17 mbar
1430 Still = 0.38 mbar, Tower = 103 mK. Close V14.
1441 Tower = 67.1 mK and falling. P1 = 0.050 mbar. 1K pot = 1.49 K / 5 mbar.
1445 The 1K pot keeps running dry and the still pressure is low. Let's try a little Still heating to see if that improves 1K pot stability. Set Still heater to 0.5 mW. P1 → 0.06 mbar, P2 = 5.0 mbar
1450 Set Still heater = 1 mW. P1 → 0.065 mbar, P2 = 5.0 mbar
1450 Set Still heater = 2 mW. P1 → 0.085 mbar, P2 = 5.0 mbar
1455 Tower = 48.0 mK and continues to cool.
1456 Set Still heater = 3 mW. P1 → 0.10 mbar, P2 = 5.0 mbar
1500 Tower = 37.6 mK
1501 Set Still heater = 3.5 mW. P1 → 0.11 mbar, P2 = 5.0 mbar. Well, this is the sort of Still pressure we are used to running at. P2 hasn't increased, but it doesn't seem to be falling yet either.
1506 Tower = 34.6 mK, Still = 1.44 K / 0.11 mbar, G1/G2 = 172/165 mbar, 1K Pot = 1.48 K / 4.5 mbar
1515 Commence LHe transfer (MB=28.8%). Don't precool transfer tubes, just wait ~15 seconds for air to be pushed out when inserting into the bath and dewar.
1555 Since starting LHe transfer, Tower temp has risen from 33.6 mK to 73.4 mK. Perhaps due to not precooling the transfer tubes.
1518 Cease LHe transfer (MB=90.32%). Tower = 107 mK, P1=0.11 mbar, P2 = 5.0 mbar
1640 Tower temp continues to increase to 119 mK. Try reducing Still heating to 2 mW. P1→0.085 mbar, but Tower temp continues to increase. Return Still heater to 3.5 mW.
1656 Since we started Still heating at 1445, G1/G2 have been rising steadily to 442/410 mbar. This is a bit high.
1700 1K pot = 1.45 K / 5 mbar. Open 1K pot needle valve +10 degrees. 1K Pot → 1.56 K / 13 mbar.
1715 Tower temp = 132.4 mK, G1/G2 flat at 443/411 mbar
1728 Fill LN2 cold trap partially.
1735 Start logging pressures in vuk-01:i\K100\20180611_pressure_log.txt
2018 Tower = 94.1 mK, G1/G2 = 389/365 mbar, P1 = 0.95 mbar, P2 = 1.8 mbar
2250 Tower = 103 mK, G1/G2 = 409/380 mbar, P1 = 0.95 mbar, P2 = 1.9 mbar

Next Day