| | this is much slower than May 16, that's good |
| 1000 | Open 1K Pot needle valve about 30 degrees. 1K Pot: 6.5 K / 0.09 mbar → 3.0 K / 7.5 mbar |
| 1005 | Open 1K Pot needle valve about 10 degrees more. 1K Pot: 2.9 K / 7.5 mbar → 2.5 K / 18 mbar |
| | MC(Cernox) has fallen from 4.6 to 3.3 K. G1(fridge) has fallen from 540 to 430 mbar. |
| 1010 | Open 1K Pot needle valve about 10 degrees more. 1K Pot: 2.5 K / 18 mbar → 2.1 K / 32 mbar |
| 1020 | Tower = 2.97 K and falling, G1(Fridge) = 293 mbar and falling |
| 1020 | Open 1K Pot needle valve about 10 degrees more. 1K Pot: 2.1 K / 32 mbar → 1.7 K / 130 mbar. After a few minutes pressure falls. |
| 1030 | Tower = 2.54 K, G1(Fridge) = 152 mbar, 1K Pot = 1.69 K / 55 mbar |
| 1040 | Tower = 2.15 K, G1(Fridge) = 72 mbar, 1K Pot = 1.61 K / 20 mbar |
| 1040 | Open 1K Pot needle valve about 10 degrees more. 1K Pot → 1.63 K / 28 mbar. |
| 1045 | Tower = 2.10 K, G1(Fridge) = 63 mbar, 1K Pot = 1.62 K / 25 mbar … then 1K Pot pressure rises off-scale, 1.77 K. Maybe we hit the lambda point. Fridge and tower had flattened out at this point. |
| 1050 | 1K Pot = 1.77 K / 1.3/30 bar (off-scale on P2) |
| 1055 | Pot pressure still off-scale, tower and fridge flat. Try closing needle valve a little. |
| 1100 | Close 1K Pot needle valve about 10 degrees less. 1K Pot → 1.75 K / 1.3/30 bar. |
| 1105 | Close 1K Pot needle valve about 10 degrees less. 1K Pot → 1.70 K / 120 mbar. |
| 1110 | 1K Pot = 1.71 K / 140±20 mbar (oscillating) |
| 1115 | Close needle valve, then switch from scroll pump to 4He pump. Open 4He exhaust to recovery (both valves) |
| 1120 | With needle valve still closed, 1K Pot → 1.37 K / 1.7 mbar, and Tower and Fridge start falling. Leave it like this for a bit. |
| 1125 | Tower = 1.98↓ K, G1(Fridge) = 54↓ mbar, 1K Pot = 1.42 K / 3.5 mbar |
| 1130 | Tower = 1.83↓ K, G1(Fridge) = 46↓ mbar, 1K Pot = 1.36 K / 1.6 mbar |
| | there was a brief increase in the pot pressure&temperature, and the fall of the fridge and tower accelerated. Let's try opening the needle valve a little. |
| 1135 | Open 1K Pot needle valve 20 degrees. 1K Pot: 1.36 K / 1.5 mbar → 1.42 K / 3.2 mbar |
| 1140 | Tower = 1.75↓ K, G1(Fridge) = 40↓ mbar, 1K Pot = 1.42 K / 3.2 mbar. Tower/Fridge cooling slowing down. |
| 1155 | Tower = 1.69↓ K, G1(Fridge) = 35↓ mbar, 1K Pot = 1.48±0.02 K / 5±1 mbar. |
| 1215 | Tower = 1.66 K, G1(Fridge) = 32 mbar, 1K Pot = 1.48±0.02 K / 5±1 mbar. Flattening out, will start circulation soon |
| 1240 | Tower = 1.65 K, G1(Fridge) = 30 mbar, 1K Pot = 1.48±0.01 K / 5.0±0.5 mbar. |
| 1240 | Open 1K Pot needle valve 10 degrees more. 1K Pot → 1.54±0.02 K / 10±2 mbar |
| 1245 | Close V9 leaving 36.9 mbar in dump. Close V3. Turn on 3He pump, which will push a little more mix into condenser. Fridge pressures rise a little. |
| 1250 | Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 90 mbar. Slight 1K Pot pressure spike to 20 mbar, then falls. Tower falls to 1.45 K. |
| 1300 | V6 → 13.2%. Condenser peaks at 90 mbar. Tower → 1.41 K. P1 falls into range. |
| 1305 | V6 → 13.5%. Condenser peaks at 105 mbar. Tower → 1.23 K. Still → 8 mbar. |
| 1310 | V6 → 13.8%. Condenser peaks at 90 mbar. Tower → 1.04 K. Still → 4.5 mbar. |
| 1315 | V6 → 14.2%. Condenser peaks at 100 mbar. Tower → 0.86 K. Still → 2.8 mbar. |
| 1320 | Tower/MC(H35)/Coldplate/Still = 0.84/0.80/0.76/1.23 K |
| 1325 | V6 → 15.0%. Condenser peaks at 100 mbar. Tower → 0.44 K. Still → 0.95 mbar. |
| 1330 | V6 → 100%. Condenser peaks at 100 mbar. Tower → 0.16 K. Still → 0.20 mbar. |
| 1345 | Tower/MC/Coldplate = 80.0/59.1/188 mK, Still = 1.15 K / 0.120 mbar, G1/G2 = 69/86 mbar, 1K Pot = 1.54 K / 8 mbar. |
| 1350 | Open V14 to let the rest of the mixture from the trap into circulation. Condenser peaks at 80 mbar, Still peaks at 20 mbar, MC rises to 73 mK and plateaus, starts falling again after a couple of minutes. |
| 1400 | Tower = 59.3 mK, Still = 3.6 mbar, G1/G2 = 55/75 mbar, 1K Pot = 1.54 K / 8 mbar. |
| 1410 | Tower = 58.0 mK, Still = 1.0 mbar, G1/G2 = 58/73 mbar, 1K Pot = 1.54 K / 8.5 mbar. |
| 1420 | Tower = 55.0 mK, Still = 0.55 mbar, G1/G2 = 60/75 mbar, 1K Pot = 1.53 K / 7.5 mbar. |
| 1420 | Close V14 |
| 1500 | Tower/MC/Coldplate = 53.1/32.6/183 mK, Still = 1.19 K / 0.110 mbar, G1/G2 = 61/77 mbar, 1K Pot = 1.54 K / 8 mbar. |
| | Tower temperature has flattened out at 53 mK |
| 1505 | Try setting Still heater to 1 mW. Condenser rises a little, to 0.12 mbar; circulation pressures rise a little. No change in tower temperature. |
| 1510 | Try setting Still heater to 2 mW. Condenser → 0.14 mbar, G1/G2 → 66/84 mbar. Tower cools very slightly, 53.0→52.5 mK. |
| 1520 | Try setting Still heater to 4 mW. Condenser → 0.16 mbar, G1/G2 → 69/89 mbar. Tower peaks to 54 mK then cools a bit, →52.2 mK. |
| 1525 | MC = 31.7 mK |
| 1600 | Messed with still heater a little more, then turned it off. |
| 1630 | Tower/MC/Coldplate = 52.2/32.1/182 mK, Still = 1.23 K / 0.100 mbar, G1/G2 = 62/77 mbar, 1K Pot = 1.54 K / 8±0.5 mbar. |
| | not cooling much now but maybe it will slowly go down over the weekend. It's a bit colder now than the first day of Run 62, and in that run it cooled to below 35 mK in a couple weeks. |
| 1655 | Fill LN2 cold trap bath |
| 2100 | Tower = 51.3 mK |