2018/05/15 (Tuesday)
| Status: Transferred LN2 yesterday. Still and MC heaters on @ 20 mW (oops) | |
| 0850 | IVC = 1.6 mbar. Start pumping out IVC with scroll, then turbo pump |
| 0850 | Turn off MC and Still heaters |
| 0955 | Fill LN2 cold trap bath |
| 0855 | MC(Cernox)/Sorb = 77.2/122.8 K, Fridge(G1) = 589 mbar |
| 1020 | Try Mini-throughput check: close off Trap 2 and V3. Only V1, V9, condenser speedivalve, still gate valve open. Turn on 3He pump, then open V6 to 12.8%. This pumps mix out of the still side and we can see if the pressure goes down on the condenser side. G1 (condenser) starts out at 590 mbar, then falls steadily at 41 mbar/min. |
| 1025 | IVC leak = 1090 pbarL/s, New level resistor = 2.982 kΩ |
| 1025 | Start blowing out LN2 |
| 1030 | IVC leak = 1050 pbarL/s, Main bath = 455 mbar above 1 atm, New level resistor = 2.975 kΩ |
| 1035 | IVC leak = 964 pbarL/s, Main bath = 434 mbar above 1 atm, New level resistor = 2.970 kΩ |
| 1038 | 77 K electrical check looks good. |
| 1040 | IVC leak = 912 pbarL/s, Main bath = 431 mbar above 1 atm, New level resistor = 2.970 kΩ |
| 1045 | IVC leak = 870 pbarL/s, Main bath = 427 mbar above 1 atm, New level resistor = 2.971 kΩ |
| 1050 | IVC leak = 826 pbarL/s, Main bath = 429 mbar above 1 atm, New level resistor = 2.973 kΩ |
| 1055 | IVC leak = 809 pbarL/s, Main bath = 426 mbar above 1 atm, New level resistor = 2.974 kΩ |
| 1100 | IVC leak = 784 pbarL/s, Main bath = 427 mbar above 1 atm, New level resistor = 2.976 kΩ |
| Throughput test procedure, Cold Trap 2: |
- All mixture in dump, LN2 traps cold, condenser & still valves open (3He pump on - optional)
- Open V9, V13b, V12b+speedivalve, V3. Wait for G1 to be close to G2.
- Close V12b+speedivalve and V3, open V1
- monitor G1
| 1510 | Perform mixture throughput test at MC = 77 K using above procedure with 3He pump on. Let G1 rise to 565 mbar before Step 3, then it falls to 64 mbar and starts to rise. Throughput = 45.4 mbar/min. |
|---|---|
| 1105 | IVC leak = 763 pbarL/s, Main bath = 378 mbar above 1 atm, New level resistor = 2.983 kΩ |
| 1110 | IVC leak = 770 pbarL/s, Main bath = 159 mbar above 1 atm, New level resistor = 2.955 kΩ |
| 1114 | Stop pressurizing main bath |
| 1115 | IVC leak = 787 pbarL/s, New level resistor = 2.948 kΩ |
| 1120 | Pump out main bath with scroll pump. Falls below 1 mbar in about 10 minutes |
| 1135 | Check pressure in warm LN2 Trap 1: 16.2 mbar. Pump out with scroll pump. |
| 1140 | Let mixture into fridge from dump through cold LN2 Trap 2. Turn off 3He pump. |
| 1145 | Pressurize MB with helium to 190 mbar above 1 atm. |
| 1145 | Check pressure in 1K Pot: 110 mbar. Pump out with scroll pump for several minutes, down to 0.11 mbar. Reducing the pressure of the 1K Pot caused a sudden decrease in Sorb temperature, 125→113 K |
| 1155 | IVC leak = 640 pbarL/s |
| 1200 | Let 63 μbar He exchange gas into IVC. |
| 1205 | MC(Cernox) = 77.3 K, Sorb = 119.7 K, IVC = 63 μbar, G1(fridge) = 596 mbar. |
| 1220 | Begin initial LHe transfer |
| 1225 | MC(Cernox) = 77.6 K, Sorb = 131.3 K, IVC = 63 μbar, G1(fridge) = 594 mbar. |
| 1235 | MC(Cernox) = 77.0 K, Sorb = 130.1 K, IVC = 30 μbar, G1(fridge) = 593 mbar. |
| 1300 | MC(Cernox) = 60.6 K, Sorb = 65.9 K, IVC = 12 μbar, G1(fridge) = 558 mbar, MB = -0.5% |
| 1310 | MC(Cernox) = 54.8 K, Sorb = 27.7 K, IVC = 10 μbar, G1(fridge) = 518 mbar, MB = 4.7% |
| 1320 | MC(Cernox) = 49.3 K, Sorb = 7.8 K, IVC = 8.6 μbar, G1(fridge) = 473 mbar, MB = 10.3% |
| 1330 | MC(Cernox) = 44.2 K, Sorb = 7.5 K, IVC = 7.4 μbar, G1(fridge) = 465 mbar, MB = 17.2% |
| 1340 | MC(Cernox) = 37.9 K, Sorb = 7.1 K, IVC = 7.4 μbar, G1(fridge) = 460 mbar, MB = 27.2% |
| 1350 | MC(Cernox) = 30.5 K, Sorb = 6.7 K, IVC = 6.3 μbar, G1(fridge) = 457 mbar, MB = 36.4% |
| 1400 | MC(Cernox) = 19.3 K, Sorb = 6.0 K, IVC = 1.9 μbar, G1(fridge) = 453 mbar, MB = 45.4% |
| 1405 | MC(Cernox) = 10.0 K, IVC = 0.74 μbar Start pumping out IVC with turbo pump |
| 1410 | MC(Cernox) = 4.7 K, Sorb = 5.5 K, G1(fridge) = 442 mbar, MB = 54.7% |
| 1415 | Fill LN2 cold trap bath |
| 1420 | MC(Cernox) = 4.5 K, Sorb = 5.6 K, G1(fridge) = 440 mbar, MB = 64.0% |
| 1430 | MC(Cernox) = 4.4 K, Sorb = 5.8 K, G1(fridge) = 439 mbar, MB = 73.0% |
| 1440 | MC(Cernox) = 4.4 K, Sorb = 5.9 K, G1(fridge) = 440 mbar, MB = 77.4% |
| 1443 | End LHe transfer @ MB = 71.3% |
| 1525 | Turbo on IVC = 1.6 nbar |