2018/05/14 (Monday)

Previous Day

Status: Discovered blockage at 77K on Saturday. Warmed up since then. Helium in IVC. Pumping out cryostat vacuum jacket overnight. Still and MC heaters on @ 20 mW
0900 MC(Cernox) = 280.6 K
0905 Still = 0.85 mbar. Still gate valve was open to P1 during the warmup
0910 Check condenser pressure, not much. Pump out both sides of fridge with scroll pump
0910 Turbo on cryostat vacuum jacket = 8.6 nbar; cryostat vacuum jacket = 100 nbar with pump valved out. Stop pumping out jacket.
0915 Check pressure in warm LN2 Trap 2: 8.4 mbar. Pump out with scroll pump
0915 Check pressure in warm LN2 Trap 1: 9.8 mbar. Pump out with scroll pump
Throughput test procedure, Cold Trap 2:
  1. All mixture in dump, LN2 traps cold, condenser & still valves open (3He pump on - optional)
  2. Open V9, V13b, V12b+speedivalve, V3. Wait for G1 to be close to G2.
  3. Close V12b+speedivalve and V3, open V1
  4. monitor G1
0940 Perform mixture throughput test at MC = 281 K using above procedure with 3He pump off. Let G1 rise to 546 mbar before Step 3, then it falls to 39 mbar and rises steadily. Throughput = 6.4 mbar/min. Compare to September, it's a little lower this time. But pressure differential is lower too.
0950 Pump mixture out of fridge to dump
1200 Let 6.3 mbar He exchange gas into IVC
1405 Check pressure in 1K Pot: 24 mbar. Pump out with scroll pump.
1430 Pump out main bath to <1 mbar
1435 Pressurize main bath with He to 301 mbar above 1 atm
1445 Let mixture into fridge through cold LN2 Trap 2. Using all of the mixture: V9 is open. 3He pump off.
1445 MC(Cernox) = 281.5 K, Sorb = >249 K (max), IVC = 6.3 mbar, G1 (fridge) = 599 mbar, Old Level resistor = 0.990 kΩ, New level resistor = 2.143 kΩ
1450 Begin LN2 transfer
1510 MC(Cernox) = 274.1 K, IVC = 5.4 mbar, G1 (fridge) = 598 mbar, Old Level resistor = 1.099 kΩ, New level resistor = 2.219 kΩ
1530 MC(Cernox) = 236.0 K, IVC = 3.0 mbar, G1 (fridge) = 595 mbar, Old Level resistor = 1.207 kΩ, New level resistor = 2.843 kΩ
1605 MC(Cernox) = 165.9 K, IVC = 2.2 mbar, G1 (fridge) = 593 mbar, Old Level resistor = 1.300 kΩ, New level resistor = 2.967 kΩ
1620 MC(Cernox) = 148.6 K, IVC = 2.2 mbar, G1 (fridge) = 591 mbar, Old Level resistor = 1.320 kΩ, New level resistor = 2.991 kΩ
1630 MC(Cernox)/Sorb = 137.4/159.2 K, IVC = 2.2 mbar, G1 (fridge) = 590 mbar, Old Level resistor = 1.334 kΩ, New level resistor = 3.009 kΩ
1645 MC(Cernox)/Sorb = 127.5/142.8 K, IVC = 1.9 mbar, G1 (fridge) = 589 mbar, Old Level resistor = 1.337 kΩ, New level resistor = 3.018 kΩ
1650 MC(Cernox)/Sorb = 122.6/136.4 K, IVC = 1.9 mbar, G1 (fridge) = 589 mbar, Old Level resistor = 1.336 kΩ, New level resistor = 3.018 kΩ
1655 Stop LN2 transfer
Cryocleaning mixture
1000 Start cleaning mixture through cold LN2 Trap 2, V6 = 12.8%
1005 Increase V6 to 13.0% to speed up circulation
1020 Increase V6 to 13.1% to speed up circulation
1050 Stop cryocleaning, Trap 2 50 minutes.
1105 Start cleaning mixture through cold LN2 Trap 1, V6 = 13.1%
1110 Increase V6 to 13.2% to speed up circulation
1155 Stop cryocleaning, Trap 1 50 minutes.
1205 Check pressure in warm Trap 2: 12.0 mbar. Pump out with scroll pump
1215 Start cleaning mixture through cold LN2 Trap 2, V6 = 12.9%
1235 Increase V6 to 13.0% to speed up circulation
1310 Stop cryocleaning, Trap 2 55 minutes.
1325 Check pressure in warm Trap 1: 12.2 mbar. Pump out with scroll pump
1330 Start cleaning mixture through cold LN2 Trap 1, V6 = 13.1%
1340 Increase V6 to 13.2% to speed up circulation
1345 Increase V6 to 13.3% to speed up circulation
1415 Stop cryocleaning, Trap 1 45 minutes.
1425 Check pressure in warm Trap 2: 4.5 mbar. Pump out with scroll pump

Next Day