| 1235 | Tighten screws of upper two seals of IVC. They can be tightened a little. |
| 1240 | Remove lower piece of IVC to redo seal. |
| 1345 | IVC screws mostly tightened. Begin pumping IVC with scroll pump. |
| 1400 | Start pumping IVC with turbo pump |
| 1430 | Turbo = 220 nbar. IVC = 14 μbar with pump valved out for 5 minutes. |
| 1430 | Start leak detector on IVC. Background about 22 pbarL/s. |
| 1435 | Squirt-test, no leak seen at IVC indium seals. Around 20 pbarL/s increase when squirting around the stripline tube elbow, particularly the KF flange just before the elbow. |
| 1455 | Check pressure in fridge lines: <0.3 mbar. Pump out with scroll pump. |
| 1530 | Raise main bath |
| 1535 | Check pressure in 1K Pot: P2 = 24 mbar. Pump out with scroll pump. |
| 1545 | Turbo = 86 nbar. IVC = 12 μbar with pump valved out for 5 minutes. |
| 1600 | Pump out main bath with scroll pump to <1 mbar on P2. |
| 1605 | With leak detector on IVC, let mixture from pump outlet through Trap 2 into fridge lines, both sides. Leak steady at 22 pbarL/s. So no leak from circulation lines to IVC. |
| 1610 | Pressurize main bath with helium to 323 mbar above 1atm on G3. After fully pressurized, IVC leak signal begins to slowly rise, 22→35 pbarL/s. Probably leaking out the top of the bath. |
| 1610 | Open v9 so that mixture in dump is also open to Trap 2 and fridge. |
| 1615 | Let 5.4 mbar He exchange gas into IVC |
| 1615 | Begin transferring LN2 |
| 1615 | MC(Cernox) = 289.5 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 594 mbar, Level resistor = 0.989 kΩ. |
| 1620 | Pause transfer out of concern that the hose might be clogged with ice. Let it warm up and dry out. |
| 1630 | Restart LN2 transfer |
| 1635 | Again transfer seems slow somehow. Dewar was just filled and is frosty, maybe it needs to be warmed up at the valve and dried out. Stop transfer again to warm things up. |
| 1645 | Restart LN2 transfer |
| 1700 | Dewar valve is completely open but LN2 transfer seems very slow. (We filled the cold trap bath with it earlier, so it worked ok then.) Hose is getting stiff and frosting up, and Tower temperature is slowly falling, but level resistor reading hasn't budged yet. |
| 1700 | MC(Cernox) = 285.2 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 596 mbar, Level resistor = 0.992 kΩ. |
| 1710 | MC(Cernox) = 282.8 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 596 mbar, Level resistor = 0.999 kΩ. |
| 1715 | MC(Cernox) = 280.0 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 596 mbar, Level resistor = 1.003 kΩ. |
| 1720 | MC(Cernox) = 276.6 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 596 mbar, Level resistor = 1.006 kΩ. |
| 1725 | MC(Cernox) = 271.5 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 596 mbar, Level resistor = 1.010 kΩ. |
| 1745 | MC(Cernox) = 246.4 K, Sorb = >249 K (max), IVC = 3.4 mbar, G1 (fridge) = 596 mbar, Level resistor = 1.024 kΩ. |
| 1810 | MC(Cernox) = 210.2 K, Sorb = >249 K (max), IVC = 3.0 mbar, G1 (fridge) = 596 mbar, Level resistor = 1.052 kΩ. |
| 1830 | MC(Cernox) = 183.9 K, Level resistor = 1.092 kΩ. |
| 1900 | MC(Cernox) = 144.7 K, Level resistor = 1.206 kΩ. |
| 1930 | MC(Cernox) = 118.4 K, Level resistor = 1.280 kΩ. |
| 1945 | MC(Cernox) = 109.7 K, Level resistor = 1.300 kΩ. |
| 2000 | MC(Cernox) = 102.7 K, Level resistor = 1.312 kΩ. |
| 2015 | MC(Cernox) = 97.1 K, Level resistor = 1.320 kΩ. |
| 2030 | MC(Cernox) = 92.4 K, Level resistor = 1.325 kΩ. |
| 2045 | MC(Cernox) = 89.0 K, Level resistor = 1.329 kΩ. |
| 2100 | MC(Cernox) = 86.1 K, Level resistor = 1.332 kΩ. |
| 2115 | MC(Cernox) = 83.9 K, Level resistor = 1.334 kΩ. |
| 2130 | MC(Cernox) = 82.2 K, Level resistor = 1.334 kΩ. |
| 2145 | MC(Cernox) = 80.9 K, Level resistor = 1.335 kΩ. |
| 2200 | MC(Cernox) = 79.9 K, Level resistor = 1.336 kΩ. |
| 2315 | MC(Cernox) = 77.8 K, Sorb = 119.2 K, IVC = 1.4 mbar, G1 (fridge) = 589 mbar, Level resistor = 1.337 kΩ. |
| 2317 | Stop LN2 transfer. Put ping pong ball on main bath outlet. |