UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20180331

2018/03/31 (Saturday)

Previous Day

Setup: DIB1/2 → SL 43/41 → 3.2 MHz filters → cables 3/1 → DCRC50(3)/55(1).
Overnight: Flash and take a set of -8V Ba series. Looks like the fridge temperature started increasing around 5:00 am. Reached a temp of around 1.3 K.
0950 Start Sequence of 8 45 min Ba series using flashandtake_R67.msl, with DCRC bias = 6 V, with 30s flash (including initial flash), 15min cooldown, 1000 beginning randoms, and a 60/50/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering.

This sequence was started after the fridge failure, with the Tower above 1 K. So this is probably junk data. Or, we could call it interesting high-temperature data. Aborted at 14:33, 5 of 8 series recorded (5th one is only 23 minutes).
1300 Fill LN2 cold trap full.
1306 1K pot pressure was very low, 1.3 mbar. Tweak needle valve a few degrees. Pressure went off scale - close it a few degrees. Settles at around 12 mbar.
1312 Stop data sequence. (actually no, it was still going)
1318 Open V14 briefly. Pressures go up, close V14, then settle. Tower = 1.01 K, G1/G2 = 360/340 mbar, P1 = 0.06 mbar, P2 = 9 mbar.
1327 Open V14 again briefly - close it. Tower = 864 mK, G1/G2 = 470/440 mbar, P1 = 0.110 mbar, P2 = 24 mbar.
1330 Start LHe transfer (MB = 32.8%)
1341 Open and close V14
1351 Tower = 768 mK, G1/G2 = 195/178 mbar, P1 = 0.065 mbar, P2 = 12 mbar.
1413 Tower = 621 mK, G1/G2 = 113/102 mbar, P1 = 0.040 mbar, P2 = 7 mbar.
1417 End LHe transfer (MB = 91.3%)
1418 Open and close V14
1424 Tower = 545 mK, G1/G2 = 255/236 mbar, P1 = 0.075 mbar, P2 = 5 mbar. Opening V14 doesn't seem to have much affect on P2 any more. May have to tweak the needle valve again?
1430 Stop data sequence. Then set bias to 0 V. (Not sure why the sequence didn't stop).
1432 Tweak 1K pot needle valve open a few degrees. P2 rises off scale, then down to around 50 mbar, fluctuates up again to around 300 mbar, and settles around 60 mbar.
1439 Tower = 307 mK, G1/G2 = 174/158 mbar, P1 = 0.075 mbar, P2 = 55 mbar.
1453 P2 falls to 24 mbar. Tweak 1K pot needle valve open a few degrees again and wait to settle.
1500 Tower = 162 mK, G1/G2 = 154/139 mbar, P1 = 0.065 mbar, P2 = 55 mbar. (I expect P2 to continue falling, not very rapidly but enough to be concerning. Send me a message if you need me to stop back in today. )
1916 Tower = 57.6 mK
1917 Start Sequence of 8 45 min Ba series using flashandtake_R67.msl, with DCRC bias = 6 V, with 30s flash(including initial flash), 15min cooldown, 1000 beginning randoms, and a 60/50/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering
2240 1K Pot pressure down to 2.2 mbar, condenser 90 mbar, still 0.04 mbar - these are all very low. Open needle valve more, also open V14 until condenser rises to 500 mbar, then close again. Tower was around 80 mK, rises to 190 mK then cools again.
2250 Condenser falling, now 266 mbar. Open V14 again, leave open a while - still pressure falling as the rest of the mixture in the dump returns to circulation. Tower increases again, to 120 mK.
2300 Still down to 4 mbar. Close V14.
2300 1K Pot down to 20 mbar. Tweak needle valve open some more.
2305 Still has fallen to 0.075 mbar, low, maybe it'll rise again. Tower 63 mK and cooling. 1K Pot 45 mbar and settling.

Next Day