2018/03/30 (Friday)

Previous Day

Setup: DIB1/2 → SL 43/41 → 3.2 MHz filters → cables 3/1 → DCRC50(3)/55(1).
Overnight: Flash and take a set of 8V series. Last flash was at 0023 this morning. Tower01 is currently crashed though…
We ran a flash and take sequence last night of 8, 8V series. Data looks good in the first 5 series, but dumps are empty starting with dump 35 of 07180329_2136. MIDAS error messages indicate that towerfe3_01 crashed around 2211
0940 Tower/MC/Coldplate = 36.2/32.2/191 mK, Still = 1.34 K / 0.085 mbar, G1/G2 = 138/125 mbar, 1K Pot = 1.38 K / 2.2 mbar. Main bath = 70.9%
0945 Tweak 1K Pot needle valve open a few more degrees. 1K Pot spikes to → 1.75 K / >400 mbar in a few seconds! Tweak 1K Pot needle valve closed a few degrees. 1K Pot oscillates around → 1.58 K / 12 mbar
0949 Fill LN2 cold trap.
1006 Return 133Ba source to safe.
1008 Restart DCRC programs.
1009 1K Pot → 1.52 K / 7 mbar
1010 Tweak 1K Pot needle valve open a few more degrees. 1K Pot → 1.57 K / 13 mbar
1031:35 Start 68 min series with initial DCRC bias = 0 V, 1000 beginning randoms, and a 70/60/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering
1034:20 Set DCRC bias = 8V using set_DCRC_bias.msl
1139 Stop run
1140:59 Start 30 min series with initial DCRC bias = 8 V, 1000 beginning randoms, and a 70/60/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering
1155 Set DCRC bias = 0V using set_DCRC_bias.msl
1225 Look for breakdown. Step up bias from 0 to 5 to 10 to 13. At 13 we saw breakdown. Then stepped back, 11 to 12 to 13. 11 and 12 showed no breakdown; 13 this time showed some hairy noise, partial breakdown. Then back to 10 then to 13. This time 13 showed no breakdown at all. Could be that after a certain time biased, the detector is less susceptible to breakdown.
1225 1K Pot fell to 4 mbar, tweak needle valve again. Settles around 20 mbar but falls again.
1230 Ground detector and flash for 30 s
1310 1K Pot has fallen to 5 mbar, tweak needle valve again. Settles around 26 mbar.
1350 Dial up MC heater to 100 uW over about half an hour. Tower settles at about 130 mK
1352 Start 35 min series at 8 V, Tower = 124-135 mK, 1000 beginning randoms, and a 70/60/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering.
1415 1K Pot has fallen to 5.5 mbar, tweak needle valve again. Settles around 20 mbar.
1415 Pot tweak caused a cooling trend, Tower = 134 mK but falling
1425 Tower = 124 mK
1430 Stop data, ground detector, flash for 30 seconds
1440 Tower falls to 127 mK then starts rising
1514 Tower = 137 mK
1514 Start 48 min series at 8 V, Tower = 138-126 mK, 1000 beginning randoms, and a 70/60/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering.
1515 1K Pot has fallen to 5.0 mbar, tweak needle valve again. Settles around 25 mbar.
1605 Stop data, ground detector
1605 Turn off MC heater (was at 100 uW)
1610 1K Pot has fallen to 9.5 mbar, tweak needle valve again. Settles around 25 mbar.
1620 Tower = 42.0 mK
1625 Flash detector for 10 seconds
1640 Look for breakdown again, with reasonably well-neutralized detector. Try 13 V and -13 V. Actually doesn't seem to break down in either case.
1645 Set bias to 10V. Will leave for 10 minutes or so to get the detector out of its early state.
1705 Detector has been at 10V for 20 min. Try 13 V : no breakdown. Go to 0V, then -13 V : no breakdown.
1713 Remove 133Ba fro safe and place in castle near fridge.
1714 Start Sequence of 8 45 min Ba series using flashandtake_R67.msl, with DCRC bias = -8 V, with 30s flash, 15min cooldown, 1000 beginning randoms, and a 60/50/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering
1715 Fill LN2 cold trap
1716 1K Pot has fallen to 9.5 mbar, tweak needle valve again.