2018/03/29 (Thursday)

Previous Day

Setup: DIB1/2 → SL 43/41 → 3.2 MHz filters → cables 3/1 → DCRC50(3)/55(1).
0352 Stop data. Set biases to 0 V. Turn on LEDs to start a bake.
0806 Turn off LEDs to end bake (4 hr 14 min)
0905 Tower/MC/Coldplate = 36.6/32.2/194 mK, Still = 1.51 K / 0.13 mbar, G1/G2 = 185/170 mbar, 1K Pot = 1.64 K / 30 mbar. Main bath = 35.6%
plan: take a series around 8 V for a while. Then standard flash, then another identical series. Want to verify that standard flash is just as good as long bake.
0917 Start 63 min series , with DCRC bias = 8 V, 1000 beginning randoms, and a 60/50/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering
0925 Don't know whether triggering is good on this data, because the copy script seems to be hanging - the data is not getting copied to /midasraw/byseries. The file /livedata/lazy_log_update.log has this line:
 09:19:32.253 2018/03/29 [Lazy_0,INFO] Starting lazy job '/home/hep/cdmssoft/scdmsDAQ/MidasDAQ/online/scripts/lazylog/copy_rawdata.py /data/chocula/k100/midasraw/livedata/RUN28432_DUMP0000.mid.gz 2>&1'
so it seems that it's stuck. Data is coming in to /livedata but it's not being copied.
The Stat1 light on DCRC1 is flickering occasionally, but on DCRC3 the brightness is constant. So could be we are triggering constantly.
1021 Fill LN2 cold trap full.
1022 Stop run, ground detector
1025 Flash for 30 s
1027 Restart MIDAS. Lazylogger starts copying over all the dumps from previous run.
Turns out this morning's series had good triggering.
1056 Start 120 min series , with DCRC bias = 8 V, 1000 beginning randoms, and a 60/50/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering
1301 Stop data, set biases to 0V, flash detector for 30 seconds
1305 Start LHe transfer (MB = 29.5%)
1355 Set biases to 6 V
plan is to look at event rate, then bring out the Ba source
1358 End LHe transfer (MB = 92.8%)
1424 Ground and flash detector.
1506 Place 133Ba source in lead castle near fridge. Source is about 8 cm from the mu-metal and 13 cm lower than the detector height.
1507 Start 45 min Ba series , with DCRC bias = 8 V, 1000 beginning randoms, and a 60/50/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering.
1530 The 1K Pot has been falling since the LHe transfer. 1K Pot = 1.54 K / 11 mbar.
1530 Tweak 1K Pot needle valve open a few more degrees. 1K Pot → 1.60 K / 24 mbar
1552 Stop data, set biases to 0V, flash for 30 s
1630 Fill LN2 cold trap full.
1632 Start Sequence of 8 45 min Ba series using flashandtake_R67.msl, with DCRC bias = 8 V, with 30s flash, 15min cooldown, 1000 beginning randoms, and a 60/50/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Good triggering
1641 1K Pot has fallen again to = 1.53 K / 9.5 mbar.
Tweak needle valve again. 1K Pot → 1.61 K / 26 mbar
1700 Looks like the 1K Pot is falling in about the same way, so I'll tweak it once more.
1K Pot: 1.59 K / 20 mbar → 1.63 K / 32 mbar