2018/03/27 (Tuesday)

Previous Day

Setup: DIB1/2 → SL 43/41 → 3.2 MHz filters → cables 3/1 → DCRC50(3)/55(1).
Bias sweep overnight. Ended with bias returned to 0 and a flash.
1005 Tower/MC/Coldplate = 36.6/30.5/187 mK, Still = 1.46 K / 0.120 mbar, G1/G2 = 167/157 mbar, 1K Pot = 1.61 K / 24 mbar. Main bath = 46.6%
1010 Fill LN2 cold trap bath.
1013 Set bias to +11 V using sequencer
1013 Start 77 min total series of +6V DCRC bias with 6 min +11V prebias manually, 1000 beginning randoms, and a 50/40/60/40 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Note, data taking includes 5min prebias time
1019 Set bias to +6 V using sequencer
1130 Set bias to 0 V using sequencer
1135 Flash for 30 seconds
1159 Start sequence of 4, 30 min series of +6V DCRC bias using takeNoflash_R67.msl, with no flash, but ground and wait 10 min between series , 1000 beginning randoms, and a 50/40/60/40 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
1203 Raise thresholds on DCRC3qI/qO from 60/40 → 70/50. Looks like we're triggering on noise.
1525 Begin LHe transfer @ MB = 38.7%
1557 End LHe transfer @ MB = 91.3%
1630 The 1K Pot temperature/pressure has been drifting downward gradually over the last few days. And it came down a little after the LHe transfer, currently 1K Pot = 1.60 K / 22 mbar.
Tweak the needle valve open a few more degrees. 1K Pot → 1.66 K / 50 mbar
1638 Start bias scan sequence of 120 min series , with DCRC bias = (3.5, 9, -3.5, -9), with 30s flash, 30min cooldown (including initial flash), 1000 beginning randoms, and a 50/40/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
1715 Raise thresholds on DCRC3qI/qO from 70/50 → 80/60. Looks like we're triggering on noise.
1716 Fill LN2 cold trap full.
1722 Raise thresholds on DCRC1qI/qO/DCRC3qI/qO from 50/40/80/60 → 60/50/90/70. Looks like we're triggering on noise.
1901 Still triggering on noise! Sequence is almost over, try disabling DCRC3QI trigger.
1907 Still triggering on noise! Also disable DCRC3QO trigger.
1914 Stop sequence after flash at 1910. Let's try sort out the triggering remotely.
1916 Restart programs.
1924 Start 0V test series with no beginning randoms, trigger DCRC1qI/qO/DCRC3qI/qO = 60/50/off/off
Triggering seems to be working fine.
1929 Start 3.5V test series with no beginning randoms, trigger DCRC1qI/qO/DCRC3qI/qO = 60/50/off/off
1934 Triggering on noise now. Disable DCRC1qO so we have: DCRC1qI/qO/DCRC3qI/qO = 60/off/off/off
1939 Still triggering on noise. Disable all triggers! DCRC1qI/qO/DCRC3qI/qO = off/off/off/off
1942 Good, no triggers. Set: DCRC1qI/qO/DCRC3qI/qO = off/off/90/off
1944 Triggering properly. Set: DCRC1qI/qO/DCRC3qI/qO = off/off/70/50
1948 Triggering properly. Set: DCRC1qI/qO/DCRC3qI/qO = 60/off/70/50
1952 Triggering properly. Set: DCRC1qI/qO/DCRC3qI/qO = 60/50/70/50
1954 Noise triggers. Set: DCRC1qI/qO/DCRC3qI/qO = 60/60/70/50
1958 Noise triggers. Set: DCRC1qI/qO/DCRC3qI/qO = 60/off/70/50
2001 Noise triggers. Set: DCRC1qI/qO/DCRC3qI/qO = off/off/70/50. (Maybe I was too quick to check the data at 1952?)
2006 Okay, triggering properly again. Seems like DCRC1 is the culprit. Set: DCRC1qI/qO/DCRC3qI/qO = 100/off/off/off
2011 Noise triggers. Set: DCRC1qI/qO/DCRC3qI/qO = 500/off/off/off
2015 Noise triggers. Set: DCRC1qI/qO/DCRC3qI/qO = 1000/off/off/off
2018 Still noise triggers. DCRC1 seems to be stuck in one of those weird trigger states. Stop run and programs then do a software RESET on DCRC1 via telnet.
2021 Start 0V test series with no beginning randoms, trigger DCRC1qI/qO/DCRC3qI/qO = 60/off/off/off
Triggering works fine.
2025 Start 3.5V test series with no beginning randoms, trigger DCRC1qI/qO/DCRC3qI/qO = 60/off/off/off
2029 Triggering properly. Set: DCRC1qI/qO/DCRC3qI/qO = 50/40/off/off
2034 Triggering properly. Set: DCRC1qI/qO/DCRC3qI/qO = 50/40/70/50 (Thresholds that we tried to start the overnight data with)
2040 Okay, it seems to be working again. Stop run, turn off bias. Not sure if there was some noise source due to the transfer that finally went away, or if DCRC1 got into a weird state and just needed a RESET. Let's try starting the sequence again.
2042 Start bias scan sequence of 120 min series , with DCRC bias = (3.5, 9, -3.5, -9), with 30s flash, 30min cooldown (including initial flash), 1000 beginning randoms, and a 50/40/70/50 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Later: All noise triggers :(
2132 Gah! Triggering on noise again. Disable DCRC1 triggers. DCRC1qI/qO/DCRC3qI/qO = off/off/70/50
2143 Still triggering on noise… I'm not sure what's going on here. Restore DCRC1 triggers and just let the thing run. Maybe it'll work itself out via magic?