2018/03/26 (Monday)

Previous Day

Setup: DIB1/2 → SL 43/41 → 3.2 MHz filters → cables 3/1 → DCRC50(3)/55(1).
0900 Tower/MC/Coldplate = 37.4/31.0/187 mK, Still = 1.45 K / 0.130 mbar, G1/G2 = 156/143 mbar, 1K Pot = 1.62 K / 26 mbar. Main bath = 30.5%
0905 Fill LN2 cold trap bath.
1004 Start sequence of 4, 30 min series of +6V DCRC bias using takeNoflash_R67.msl, with no flash, but ground and wait 30 min between series , 1000 beginning randoms, and a 50/40/60/40 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
1237 Start LHe transfer (MB = 24.4%)
1339 End LHe transfer (MB = 72.6%) Dewar ran empty.
1442 Start 60 min series of +6V DCRC bias with 5 min +12V prebias using flashandtake_wPrebias_R67.msl, with 30s flash, 30 min cooldown before series , 1000 beginning randoms, and a 50/40/60/40 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length). Note, data taking includes 5min prebias time All noise triggers [image not archived: doubt2.svg]
1659 Fill LN2 cold trap full.
1709 Start bias scan sequence of 60 min series , with DCRC bias = (5,6,7,8,9,10,11,12), with 30s flash, 30min cooldown, 1000 beginning randoms, and a 50/40/60/40 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).