start › cdms › k100 › run_diary › 20180322

2018/03/22 (Thursday)

Previous Day

Setup: DIB1/2 → SL 43/41 → 3.2 MHz filters → cables 3/1 → DCRC50(3)/55(1).
Status: Cooled to base yesterday, baked detector overnight.
0920 Tower/MC/Coldplate = 520/174/245 mK, Still = 1.42 K / 0.200 mbar, G1/G2 = 100/96 mbar, 1K Pot = 1.66 K / 40 mbar. Main bath = 22.5%
0925 End bake by turning off all four LEDs
0930 Fill LN2 cold trap bath
Minutes since end of bake Tower [K]
0 520
5 96
10 68
15 55
20 49.4
25 46.5
30 44.2
35 43.4
40 42.8
50 42.2
60 42
1035 Tower = 41.8 mK
1222 Begin LHe transfer @ 17.7%
1314 End LHe transfer @ 91.2%
At beginning and end of transfer, Tower temperature spiked a few mK. At the beginning the 1K Pot pressure spiked, and after that began to drift downward…
1320 1K Pot = 1.62 K / 30 mbar. Tweak needle valve open a little more. 1K Pot → 1.66 K / 45 mbar.
1330 Set bias on all four charge channels to 12 V. Tower spikes 42→44 mK then falls back
1334 Start 3hr 20min series of 12V data (DCRC bias), with 1000 beginning randoms, and a 80/80/100/80 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
1635 Fill LN2 trap full.
1659 Start bias scan sequence of 30 min series , with DCRC bias = (1,3,5,7,9,11,-1,-3,-5,-7,-9,-11), with 30s flash, 30min cooldown, 1000 beginning randoms, and a 80/80/100/80 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).