start › cdms › k100 › run_diary › 20180321

2018/03/21 (Wednesday)

Previous Day

Status: Initial LHe transfer yesterday. Mixture open from dump to fridge through (cold) Trap 1.
0910 Tower/Sorb/1K Pot = 5.28/8.3/6.43 K, G1(fridge) = 550 mbar, Main bath = 62.4%. DIB1 qIbias-QET D = 17.01 nS
0910 Turbo on IVC = 2.2 nbar. IVC = 16 nbar with turbo valved out.
0915 Fill LN2 cold trap bath
0920 Pump out line to 1K Pot with scroll pump. Valve out pump and open 1K Pot speedivalve to check pot pressure: 7/30 bar (on mechanical gauge), pretty high. Pot temperature drops when valve opened, to 4.6 K then back up.
0925 Pump out 1K Pot with scroll pump. Dips to 4.6 K then rises again.
0930 Switch to pumping 1K Pot with 4He pump. Open valve on pump exhaust line and valve on recovery line to connect pump exhaust to recovery.
0930 1K Pot = 6.47 K / 0.50 mbar
Open 1K Pot Needle Valve to ~10 degrees. 1K Pot → 3.2 K / 1/30 bar (mechanical gauge; off-scale on P2)
0940 1K Pot pressure oscillating strongly, falling overall, minimum around 4 mbar. Tower temperature falling.
0945 Tower/Sorb/1K Pot = 4.85/4.0/2.80 K, G1(fridge) = 473 mbar. DIB1 qIbias-QET D = 14.70 nS. Tower temperature falling rapidly.
Tower = 4.00 K, DIB1 qIbias-QET D = 11.5 nS.
0950 Tower = 3.37 K, DIB1 qIbias-QET D = 0.50 nS.
Tower = 3.35 K, DIB1 qIbias-QET D = 0.15 nS.
0950 1K Pot = 2.85 K / 4.0 mbar, flat
Open Needle Valve a few more degrees. 1K Pot → 2.77 K / 4.5 mbar
0955 Open Needle Valve a few more degrees. 1K Pot → 2.72 K / 5.0 mbar
0955 Tower = 3.27 K, DIB1 qIbias-QET D = 0.07 nS.
0958 Open Needle Valve a few more degrees. 1K Pot → 2.70 K / 6.0 mbar
1003 Open Needle Valve a few more degrees. 1K Pot → 2.40 K / 10.0 mbar
1004 Tower = 3.08 K, DIB1 qIbias-QET D = 0.07 nS.
1016 Open Needle Valve a few more degrees. 1K Pot → 1.84 K / 13.0 mbar
1018 Tower = 2.65 K, DIB1 qIbias-QET D = 0.07 nS.
1032 Open Needle Valve a few more degrees. 1K Pot → 1.56 K / 11.0 mbar, pressure starting to fall.
1033 Tower = 2.17 K, DIB1 qIbias-QET D = 0.06 nS.
1042 1K Pot → 1.51 K / 6.5 mbar, flattening out.
1044 Open Needle Valve a few more degrees, no change. 1K Pot → 1.51 K / 6.5 mbar
1045 Open Needle Valve a few more degrees, still no change. 1K Pot → 1.52 K / 6.5 mbar
1048 Tower = 2.00 K, levelling off DIB1 qIbias-QET D = 0.07 nS.
1050 Open Needle Valve a few more degrees. Got a response this time. 1K Pot → 1.68 K / 50 mbar and oscillating
1110 Tower = 1.90 K (flat), Fridge(G1) = 53.1 mbar (falling slowly), 1K Pot = 1.68 K / 50±5 mbar
1155 Tower = 1.84 K (flattish), Fridge(G1) = 47.9 mbar (falling slowly), 1K Pot = 1.68 K / 50±5 mbar
1235 Tower = 1.825 K (flattish), Fridge(G1) = 45.7 mbar (flattish), 1K Pot = 1.67 K / 40±5 mbar
1235 Close V9 leaving 35.3 mbar in dump. Close V3. Turn on 3He pump, which will push a little more mix into condenser. Fridge pressures rise a little then fall again.
1250 Tower = 1.83 K (flattish), Fridge(G1) = 46.0 mbar (flattish), 1K Pot = 1.67 K / 45±2 mbar
1255 Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 115 mbar. 1K Pot pressure spikes to 110 mbar then returns. Tower falls to 1.65 K.
1300 V6 → 13.2%. Condenser peaks at 115 mbar. Tower → 1.57 K.
1310 V6 → 13.5%. Condenser peaks at 125 mbar. Tower → 1.46 K.
1315 Still pressure falling, Still = 20 mbar
1315 V6 → 13.8%. Condenser peaks at 130 mbar. Tower → 1.15 K. Still → 7.0 mbar
1320 V6 → 14.2%. Condenser peaks at 125 mbar. Tower → 1.06 K. Still → 3.6 mbar
1325 MC(H35)/Coldplate/Still = 1.02/0.97/1.40 K
1330 Tower = 0.80 K
1330 V6 → 15.0%. Condenser peaks at 130 mbar. Tower → 560 mK. Still → 1.7 mbar
1340 Condenser/Still = 114/1.2 mbar
1340 V6 → 100%. Condenser peaks at 120 mbar. Tower → 200 mK. Still → 0.22 mbar
1350 Tower/MC(H35)/Coldplate/Still = 0.097/0.058/0.187/1.38 K
1415 Tower/MC/Coldplate = 90.7/41.7/168 mK, Still = 1.27 K / 0.130 mbar, G1/G2 = 76/76 mbar, 1K Pot = 1.67 K / 45 mbar.
1415 Open V14 to let the rest of the mixture from the trap into circulation. Condenser peaks at 105 mbar, Still peaks at 50 mbar, MC rises to 111 mK and plateaus, starts falling again after a couple of minutes.
1430 Tower = 90.5 mK, Still = 2.0 mbar, G1/G2 = 69/70 mbar, 1K Pot = 1.67 K / 45 mbar.
1500 Tower = 82.5 mK, Still = 0.4 mbar, G1/G2 = 75/74 mbar, 1K Pot = 1.67 K / 40 mbar.
1500 Close V14
1525 Tower/MC/Coldplate = 78.9/36.4/174 mK, Still = 1.42 K / 0.130 mbar, G1/G2 = 75/74 mbar, 1K Pot = 1.67 K / 45 mbar.
1535 Plug cables into DCRCs
Setup: DIB1/2 → SL 43/41 → 3.2 MHz filters → cables 3/1 → DCRC50(3)/55(1).
1540 Power up DCRCs and start MIDAS programs
Tower temperature rises 76→78 mK, then falls back to 76 mK.
1545 Set all four charge biases to +12 V. Seeing some pulses in all four channels. Setting the bias seemed to cause another short temperature spike, maybe this just happened when the first channels were biased and the others were still at ground.
1555 Adjust thresholds on individual channels
1555 Start 60min series of 12V data (DCRC bias), with 1000 beginning randoms, and a 80/80/100/80 bin threshold trigger on DCRC1qI/qO/DCRC3qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
1640 Tower = 67.8 mK, falling steadily
1655 Set bias on all four channels back to 0 V. Again this causes a temperature spike on the Tower thermometer, 64.3→67.2 mK and back.
1700 Fill LN2 cold trap bath
1705 Tower = 63.4 mK, falling steadily
1705 Turn on DCRC1 LED 1. Tower → 0.25 K.
Turn on DCRC1 LED2. Tower → 0.37 K
Turn on DCRC3 LED1. Tower → 0.45 K
Turn on DCRC3 LED2. Tower → 0.51 K
Leave detector baking overnight.
1710 Main bath = 49.7%

Next Day