Previous Day
| | Status: warming. Main bath empty. LN2 traps pumped out, mixture in dump. 1K Pot evacuated and isolated. Helium and nitrogen in cryostat vacuum jacket. |
| 0905 | MC(Cernox) = 258.7 K
Warming time constant now only 15 hours, suggests fridge will hit 273 K late this afternoon, 16:20. |
| 1100 | MC(Cernox) = 263.3 K. Should hit 273 K at about 16:07 |
| 1510 | Check the other two side coaxes that we can use this run (these are sCDMS short, with names like ST-4-something) |
| Side coax | ST-3-8 |
| | date: | 2018-03-02 | who: | MF |
| +lead | - lead | signal | typical | measured |
| D11 | T11 | Qo Rbias | 25 nS | 25 nS |
| T12 | T13 | Qo Rfeedback | 25 nS | 25 nS |
| D11 | T12 | Qo C coupling | <1 nS | 0.06 nS |
| D12 | T14 | Qi Rbias | 25 nS | 25 nS |
| T15 | T16 | Qi Rfeedback | 25 nS | 25 nS |
| D12 | T15 | Qi C coupling | <1 nS | 0.05 nS |
| all | chassis | isolation | <1 nS | OK |
The one above is the extra one, looks like it's in good working order
| Side coax | ST-3-9 |
| | date: | 2018-03-02 | who: | MF |
| +lead | - lead | signal | typical | measured |
| D11 | T11 | Qo Rbias | 25 nS | 0.05 nS |
| T12 | T13 | Qo Rfeedback | 25 nS | 0.08 nS |
| D11 | T12 | Qo C coupling | <1 nS | 12 kΩ |
| D12 | T14 | Qi Rbias | 25 nS | 0.07 nS |
| T15 | T16 | Qi Rfeedback | 25 nS | 25 nS |
| D12 | T15 | Qi C coupling | <1 nS | 0.06 nS |
| all | chassis | isolation | <1 nS | OK |
The one above is the one I tried to fix and failed. It sure is a mess. Not only 3 open resistors, but a shorted or blown capacitor.
| 1615 | Check IVC pressure: 120 mbar. Note that the overpressure valve was opened yesterday, worried that it might have trapped helium. |
| 1620 | Start letting N2 into IVC, interestingly the pressure on the Pfeiffer gauge initially went down, not up. Helium must read higher pressure than nitrogen on that gauge. |
| 1635 | Fill IVC with about 1 bar of N2 |
| 1635 | MC(Cernox) = 274.1 K by the LabVIEW vi. 272.8 K by the LakeShore bridge display. |
| 1645 | Measure pressure in cryostat vacuum jacket: 12 mbar, like yesterday. |
| 1645 | Begin pumping out cryostat vacuum jacket with scroll pump |
| 1700 | Switch to pumping out cryostat vacuum jacket with turbo pump |