2018/02/06 (Wednesday)

Previous Day

1644 Take DCRC calibration data with DCRC2 on the miniBob triggering on randoms and 2mVpp, 1 kHz square wave fed in to QI from the Agilent 33521A function generator.
1649 Data is not copying to midasraw, seems like lazylogger might be dead. Restart Midas. Seems to have worked.
1657 Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 50ADC, and 2mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator.
Great, looks like it got a mix of pulse polarities with only one pulse per trace. This should be about the signal level of the 60 keV events.
1701 Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 50ADC, and 10mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator.
1712 Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 50ADC, and 2 mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator set to High Impedance output. Failed to trigger!
1713 Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 50ADC, and 10 mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator set to High Impedance output.
1716 Take DCRC calibration data with DCRC3 on the miniBob, QI trigger = 50ADC, and 2 mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator set to High Impedance output. Short prepulse length
1719 Take DCRC calibration data with DCRC3 on the miniBob, QI trigger = 50ADC, and 2 mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator set to High Impedance output.
1721 Take DCRC calibration data with DCRC3 on the miniBob, QI trigger = 50ADC, and 10 mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator set to High Impedance output.
1725 Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 20ADC, and 2 mVpp, 5 Hz square wave fed in to QI from the Agilent 33521A function generator set to High Impedance output. Lower trigger seems to have worked.