| | DCRC calibration data taken yesterday may have been flawed. The square wave signal used as input from the Agilent signal generator does not appear to be symmetric. Will try it with the older Wavetech function generator which can't go to as low of amplitude, but appears to give a symmetric signal. |
| 1059 | Take DCRC calibration data with DCRC3 on the miniBob, QI trigger = 30ADC, and ~40 mVpp, 20 Hz ~square wave fed in to QI from the Wavetech 182A function generator. Data looks decent. |
| 1103 | Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 20ADC, and ~40 mVpp, 20 Hz ~square wave fed in to QI from the Wavetech 182A function generator. Data looks decent. |
| | Stick a couple BNC attenuators on the Wavetech function generator output so we can get down to smaller amplitudes. |
| 1135 | Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 20ADC, and ~10 mVpp, 20 Hz ~square wave fed in to QI from the Wavetech 182A function generator. Good data. |
| 1139 | Take DCRC calibration data with DCRC3 on the miniBob, QI trigger = 30ADC, and ~10 mVpp, 20 Hz ~square wave fed in to QI from the Wavetech 182A function generator. Good data. |
| 1141 | Take DCRC calibration data with DCRC3 on the miniBob, QI trigger = 30ADC, and ~2 mVpp, 20 Hz ~square wave fed in to QI from the Wavetech 182A function generator. Good data. |
| 1145 | Take DCRC calibration data with DCRC2 on the miniBob, QI trigger = 20ADC, and ~2 mVpp, 20 Hz ~square wave fed in to QI from the Wavetech 182A function generator. Good data. |