| | Setup: DIB1/3/4 → SL 27/19/20 → 3.2 MHz filters → cables 2,1,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3). DCRC4 → HV power supply. |
| 0925 | Tower/MC/Coldplate = 52.8/55.6/154 mK, MC heater = 8.0 μW, Still = 1.3 K / 0.110 mbar, G1/G2 = 244/226 mbar, 1K Pot = 1.62 K / 26 mbar, MB = 38.2% |
| 0930 | Fill LN2 cold trap bath. |
| 1100 | Having trouble processing the data from last night. Not sure if it's an issue with the raw data format or my processing settings. It's easy enough to take some more and check. |
| 1119 | Start 5 min long trace randoms transition data series at HV=0V, Driver Gains = 2x, QET biases = (430/340/380)/(400/350/430) uA, Front-End Gains = 10/5 on DCRC1/3, DCRC1,3 trace length = 13.1072ms, Trigger on 20Hz randoms |
| 1130 | Seems to process fine now. |
| 1151 | Start 5 min long trace randoms SC/transition data series at HV=0V, Driver Gains = 2x, QET biases = (0/340/380)/(400/350/430) uA, Front-End Gains = 10/5 on DCRC1/3, DCRC1,3 trace length = 13.1072ms, Trigger on 20Hz randoms |
| 1202 | Start 5 min long trace randoms SC data series at HV=0V, Driver Gains = 2x, QET biases = (0/0/0)/(0/0/0) uA, Front-End Gains = 10/5 on DCRC1/3, DCRC1,3 trace length = 13.1072ms, Trigger on 20Hz randoms |
| 1222 | Restore QET biases to (430/340/380)/(400/350/430) uA |
| 1621 | Commence LHe transfer (27.7%) |
| 1656 | End LHe transfer (89.7%) |
| 1702 | Fill LN2 cold trap full. |