| 1540 | Dial HV up to 70 V with 15% 5-minute prebias, using changeHV_0_70_preBias15_5.msl |
| 1545 | Tower = 54.7 mK, pretty stable |
| 1545 | HV = 69.86 V, V_leak = 3.28 mV |
| 1550 | Start sequence of 4 60-minute background data series at HV = 70V, Driver Gains = 2x, QET biases = (430/340/380)/(400/350/430) uA, Front-End Gains = 10/5 on DCRC1/3, 1000 initial randoms, trigger on DCRC3pD threshold 14, MC heater = 8.0 μW, Tower = 54.7 mK |
| 1559 | Reduce trigger threshold from 14 to 12. It looks like we are only triggering on strong pulses. Also it looks like we are avoiding the trigger bursts that we got before, caused by driving DCRC3pD into the strong oscillations after a muon. |
| 1605 | Reduce trigger threshold from 12 to 11. This looks good - it occasionally triggers in the noise. No trigger bursts, looks like. |
| 1721 | Fill LN2 cold trap full. |
| 2118 | Tower = 55.5 mK |
| 2119 | Dial down HV bias to 0V and flash using changeHV_70_0_thenflash.msl |
| 2203 | Tower peaked at 550mK, fell to 60mK after 10min, 56mK after 20min, and 55mK after 34min |
| 2223 | Start 5 min long trace randoms transition data series at HV=0V, Driver Gains = 2x, QET biases = (430/340/380)/(400/350/430) uA, Front-End Gains = 10/5 on DCRC1/3, DCRC1,3 trace length = 13.1072ms, Trigger on 20Hz randoms |
| 2230 | Start 5 min long trace randoms SC/transition data series at HV=0V, Driver Gains = 2x, QET biases = (0/340/380)/(400/350/430) uA, Front-End Gains = 10/5 on DCRC1/3, DCRC1,3 trace length = 13.1072ms, Trigger on 20Hz randoms |
| 2242 | Start 5 min long trace randoms SC data series at HV=0V, Driver Gains = 2x, QET biases = (0/0/0)/(0/0/0) uA, Front-End Gains = 10/5 on DCRC1/3, DCRC1,3 trace length = 13.1072ms, Trigger on 20Hz randoms |
| 2250 | Ooops! Was actually trying to take randoms at 2000 Hz! Given the readout bottleneck, the above series are probably at something like 65Hz really. |
| 2251 | Restore 55mK QET bias setting from table above. |