start › cdms › k100 › run_diary › 20171206

2017/12/06 (Wednesday)

Previous Day

Setup: DIB1/3/4 → SL 27/19/20 → 3.2 MHz filters → cables 2,1,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3). Also DCRC4 is powered and hooked up to control the HV setting.
0855 Tower/MC/Coldplate = 21.6/23.9/149 mK, Still = 1.2 K / 0.100 mbar, G1/G2 = 248/229 mbar, 1K Pot = 1.62 K / 24 mbar, MB = 33.9%
new milestone: Tower thermometer resistance is above 300 kΩ!
0900 Fill LN2 cold trap full.
0915 Increase Driver Gains to x8 and adjust Driver Offsets to get all baselines around -0.3 V.
0917 Dial HV up to 70 V with prebiasing using sequence changeHV_0_70_preBias15_5.msl
0925 HV = 69.89 V, V_leak = 3.14 mV
0945 Messing with Driver Gains and Driver Offsets and trigger threshold. A lot of channels were ADC saturating, so I've set the Driver Gains down to x2.
0953 Start 60-minute background data series at HV = 70V, Driver Gains = 2x, QET biases = 480/360/450/480/380/490 uA, Front-End Gains = 10/5 on DCRC1/3, 1000 initial randoms, trigger on DCRC3pD (Side1pF) threshold = 12
0955 HV = 69.89 V, V_leak = 2.96 mV
1030 HV = 69.87 V, V_leak = 2.77 mV
1053 Stop data series.
1055 Estimate event rate from random traces. Set PhononWaveformLength on DCRC1 to 131072 which is 52.4288 ms and counted visible pulses in 42 pA traces. Counted 70 pulses in 2.20201 seconds, so event rate is around 31.8 ± 4 Hz.
1108 Place PuBe source in lab, in its paraffin barrel centered behind lead wall behind cryostat shield.
1110 Estimate event rate from random traces. Set PhononWaveformLength on DCRC1 to 131072 which is 52.4288 ms and counted visible pulses in 24 pA traces. Counted 92 pulses in 2.25829 seconds, so event rate is around 73.1 ± 8 Hz.
1122 Start 60-minute PuBe data series at HV = 70V, Driver Gains = 2x, QET biases = 480/360/450/480/380/490 uA, Front-End Gains = 10/5 on DCRC1/3, 1000 initial randoms, trigger on DCRC3pD (Side1pF) threshold = 12
note that detector has been biased for about 2 hours without flashing. Will probably flash after this run.
DCRC3pB railed some time during the 5th dump
1125 HV = 69.87 V, V_leak = 2.69 mV
1220 HV = 69.86 V, V_leak = 2.66 mV
1222 Stop data series.
1225 Adjust driver offset on DCRC3pB. It did not fall into superconducting, but it did require an adjustment to the offset.
1230 Change Front-end Gain on DCRC3pB from 5 to 3. Maybe this will make it more stable, stay locked.
1231 Start 60-minute PuBe data series at HV = 70V, Driver Gains = 2x, QET biases = 480/360/450480/380/490 uA, Front-End Gains = 10/10/103/5/5, 1000 initial randoms, trigger on DCRC3pD (Side1pF) threshold = 12
DCRC1pC has pretty bad oscillations in the pulses, may need to reevaluate its QET bias
DCRC3pB shifted several times and was practically railed by the end of Dump 16. sucks.
1330 HV = 69.86 V, V_leak = 2.66 mV
1332 Stop data series. Detector was biased without flashing for about 4 hours.
1335 Adjust driver offset on DCRC3pB. Hoping that it becomes stable. Maybe should look at SQUID settings again.
1335 Dial HV down to zero using sequence script changeHV_70_0.msl
1405 Flash LEDs on DCRCs 1 and 3 for 30 seconds
note: after 30 minutes Tower was still about 0.4 mK above base. It reached base after about 45 minutes.
1416 Begin LHe transfer @ 26.0%
1530 Tried adjusting DCRC3pB SQUID settings but they appear to be optimal already. Restore Front-end Gain to 5.
1535 Dial HV up to 70 V with prebiasing using sequence changeHV_0_70_preBias15_5.msl
1538 Remove PuBe source from lab and put in PAN 443b storage.
1545 Tower = 21.6 mK
1550 Estimate event rate from random traces. Set PhononWaveformLength on DCRC1 to 131072 which is 52.4288 ms and counted visible pulses in 42 pA traces. Counted 65 pulses in 2.20201 seconds, so event rate is around 29.5 ± 4 Hz.
1555 Start 60-minute background data series at HV = 70V, Driver Gains = 2x, QET biases = 480/360/450480/380/490 uA, Front-End Gains = 10/10/105/5/5, 1000 initial randoms, trigger on DCRC3pD (Side1pF) threshold = 12
1612 Baseline of DCRC3pB has crept down a bit. Try adjusting Driver Offset from 0.5 to 0.4
1620 Triggering on noise a bit too often I think. Raise trigger threshold from 12 to 15.
1628 Triggering well above noise now. Lower trigger threshold from 15 to 14.
1655 Stop data series.
1704 Place PuBe source in lab, in its paraffin barrel centered behind lead wall behind cryostat shield.
1710 Fill LN2 cold trap bath
1715 Dial HV down to zero using sequence script changeHV_70_0.msl
1718 Start 60 minute PuBe data series at HV = 70 V followed by flash using sequence HVseries-70V-15_5Prebias-30Flash-30wait-1hrData.msl Driver Gains = 2x, QET biases = 480/360/450/480/380/490 uA, Front-End Gains = 10/10/10/5/5/5, 1000 initial randoms, trigger on DCRC3pD (Side1pF) threshold = 14.
2041 DCRC3PB railed low some time in the 37th dump.
2247 Change DCRCpB,PD driver offsets. Doing it remotely via the DCRC control GUI. Initially, the values displayed in the GUI were wrong, which really messed up the driver gains when I adjusted them from what I thought the current values were.
2256 Start 60 minute PuBe data series at HV = 70 V followed by flash using sequence HVseries-70V-15_5Prebias-30Flash-30wait-1hrData.msl Driver Gains = 2x, QET biases = 480/360/450/480/380/490 uA, Front-End Gains = 10/10/10/5/5/5, 1000 initial randoms, trigger on DCRC3pD (Side1pF) threshold = 14.

Next Day