2017/12/02 (Saturday)

Previous Day

Setup: DIB1/3/4 → SL 27/19/20 → 3.2 MHz filters → cables 2,1,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3). Also DCRC4 is powered and hooked up to control the HV setting.
0935 Tower = 22.4 mK
0940 Adjust driver offsets remotely.
0946 Start Noise HV scan sequence HVscan-30_50V-15_5PreBias-30Flash-5wait-1hrData.msl Sequence takes data at 30, 40, 50 V HV biases with 15% prebias for 5 min and 30s flash with 5 min wait between each 1 hr series. Random triggering, 1000 BOR randoms. DriverGains=1, QETbias = (DCRC1PA/PB/PC)/(DCRC3PB/PC/PD) = (480/370/510)/(470/380/500). Cooldown time is short to observe baseline shifts during post-flash cooldown.
0958 HV = 30.01 V, V_leak = 1.46 mV
1112 HV = 39.98 V, V_leak = 1.57 mV
1204 HV = 39.98 V, V_leak = 1.35 mV
1244 HV = 50.03 V, V_leak = 1.80 mV
1443 Fill LN2 cold trap full.
1443 Tower/MC/Coldplate = 22.5/29.2/157 mK, Still = 1.5 K/0.100 mbar, G1/G2 = 239/218 mbar, 1K Pot = 1.62 K / 24 mbar, MB = 27.4%
1455 Start LHe transfer (MB = 27.1%)
1721 End LHe transfer (MB = 89.5%)
2215 put out 22Na source and leaving out overnight. will remove at 10am tomorrow morning