start › cdms › k100 › run_diary › 20171128

2017/11/28 (Tuesday)

Previous Day

Setup: DIB1/3/4 → SL 27/19/20 → 3.2 MHz filters → cables 2,1,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3). Also DCRC4 is powered and hooked up to control the HV setting.
0905 Tower/MC/Coldplate = 23.4/25.6/153 mK, Still = 1.4 K / 0.110 mbar, G1/G2 = 250/234 mbar, 1K Pot = 1.64 K / 34 mbar, MB = 34.5%
0910 Fill LN2 cold trap full.
0943 Use PulseViewer to view traces while adjusting driver offsets with DCRCcontrol GUI.
0945 Restart MIDAS in preparation for IBIS scan.
0946 Let's prebias for 5 min. Turn on power supply with DCRC4QObias = 2V. Slowly ramp up to 52 V at 1V/sec by changing DCRC4/Charge/Bias (V)[0] by 0.1 V/sec by hand in the ODB. HV/DCRC4QI = 52.56/5.2 V
0954 Slowly ramp down to 45 V at 1V/sec by changing DCRC4/Charge/Bias (V)[0] by 0.1 V/sec by hand in the ODB. HV/DCRC4QI = 45.07/5.99 V
0956 Begin IbIs data sequence on DCRC1pC with HV=45V from 650 to 400 uA in -1.0 uA steps, 3 seconds of data per step. (IbIsDCRC1pC.msl)
Data looks OK.
1046 Sequence completed. HV = 44.97 V, V_leak = 1.83 mV
1047 Slowly ramp bias down to 0V and turn off power supply.
1048 Flash for 30s using flashandflash GUI.
1130 Tower was at 22.4 mK before flash, peaked at 476 mK, fell to 29 mK in 10 min, 24.5 mK in 20 min
1223 Tower = 22.8 mK
1227 Prebias for 5 min. Turn on power supply with DCRC4QObias = 2V. Slowly ramp up to 52 V at 1V/sec by changing DCRC4/Charge/Bias (V)[0] by 0.1 V/sec by hand in the ODB. HV/DCRC4QI = 52.51/5.2 V
1234 Slowly ramp down to 45 V at 1V/sec by changing DCRC4/Charge/Bias (V)[0] by 0.1 V/sec by hand in the ODB. HV/DCRC4QI = 45.01/5.99 V
1238 Begin IbIs data sequence on DCRC3pB with HV=45V from 750 to 400 uA in -1.0 uA steps, 3 seconds of data per step. (IbIsDCRC3pB.msl) Note:Forgot to reset MIDAS this time.
1340 Sequence is stuck at 419 uA. MIDAS web page is stuck.
1345 Kill and restart MIDAS, start DAQ programs
1350 HV = 44.98 V, V_leak = 1.66 mV
1350 Slowly ramp bias down to 0V and turn off power supply.
1355 Flash for 30s using flashandflash GUI.
1407 Start LHe transfer @ MB = 26.7%
1507 Finish LHe transfer @ MB = 90.4%
1525 Tower = 23.4 mK
1540 Adjust all six Driver Offsets to get baselines around 0.8 V
1545 Prebias for 5 min: Turn on power supply with DCRC4QObias = 2V. Slowly ramp up to 52 V at 1V/sec by changing DCRC4/Charge/Bias (V)[0] by 0.1 V/sec by hand in the ODB. HV/DCRC4QI = 52.51/5.2 V
1555 Slowly ramp down to 45 V at 1V/sec by changing DCRC4/Charge/Bias (V)[0] by 0.1 V/sec by hand in the ODB. HV/DCRC4QI = 45.02/5.99 V
We'll redo DCRC1pB IbIs because the first time we did this one at 45 V it failed.
1601 Begin IbIs data sequence on DCRC1pB with HV=45V from 600 to 300 uA in -1.0 uA steps, 3 seconds of data per step. (IbIsDCRC1pB.msl)
Series continued to completion. Data looks OK.
1935 HV = 44.99 V, V_leak = 1.53 mV
2003 HV = 44.99 V, V_leak = 1.51 mV
2004 Slowly ramp bias down to 0V and turn off power supply.
2009 Flash for 30s using flashandflash GUI.
2025 Temp monitoring webpage is suddenly showing weird temp values for the last 12 hours. Looks like ssh/sftp to vuk-01 aren't working properly right now and it's plotting old data. Not sure why.