2017/11/16 (Thursday)

Previous Day

Initial LHe transfer yesterday. Pumping IVC overnight. Mixture open from dump to fridge through (cold) Trap 2.
0920 Tower/Sorb/1K Pot = 4.66/8.2/6.36 K, G1(fridge) = 551.8 mbar, Turbo(on IVC) = 1.4 nbar, IVC = 12 nbar, MB = 72.2%
Stop pumping on IVC
0925 Fill LN2 cold trap bath
0930 Pump out line leading to 1K Pot, then open 1K Pot speedivalve. (Note that 1K Pot was evacuated before LHe transfer and needle valve was closed.) About 0.25 bar in 1K Pot. Pot temperature dips to 4.5 K when valve opened, then rises again. Pump out for a few seconds with scroll pump, then switch to 4He pump. Open the two valves to connect 4He exhaust to recovery.
0935 1K Pot = 6.38 K / 0.38 mbar
Open 1K Pot Needle Valve to ~10 degrees. 1K Pot → 3.3 K / 3.2 mbar
0940 Open Needle Valve a few more degrees. 1K Pot → 3.1 K / 4.0 mbar
Open Needle Valve a few more degrees. 1K Pot → 2.9 K / 4.5 mbar
0945 Open Needle Valve a few more degrees. 1K Pot → 2.6 K / 7.5 mbar
0950 Tower temp and Fridge pressure falling. Tower = 3.30 K, G1(fridge) = 425.0 mbar
0955 Open Needle Valve a few more degrees. 1K Pot → 2.4 K / 15 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.8 K / 30 mbar
1000 Open Needle Valve a few more degrees. 1K Pot → 1.7 K / 38 mbar
1010 Tower = 2.62 K (falling), 1K Pot = 1.66 K, Fridge(G1) = 173 mbar (falling)
1015 Tower = 2.38 K (falling), 1K Pot = 1.63 K, Fridge(G1) = 111 mbar (falling)
1020 Tower = 2.17 K (falling), 1K Pot = 1.58 K, Fridge(G1) = 75 mbar (falling)
1020 1K Pot = 1.58 K / 13 mbar, pressure steadily falling.
Open Needle Valve a few more degrees. 1K Pot → 1.58 K / 13 mbar, flattening out
1025 Open Needle Valve a few more degrees. 1K Pot → 1.62 K / 24 mbar
1030 Tower = 2.09 K (flattening), 1K Pot = 1.58 K, Fridge(G1) = 62.3 mbar (flattening)
1030 1K Pot surge, pressure on P2 goes off-scale, temperature → 1.79 K. Pressure on mechanical gauge is around 1 “Hg, which is only around 35 mbar.
1035 1K Pot pressure on P2 is still maxed out, temperature steady.
Close Needle Valve a few degrees. Pressure oscillates strongly.
1040 1K Pot = 1.73 K, 600±300 mbar. Tower temperature was flat but now begins to fall again.
1045 Tower = 2.02 K, Fridge(G1) = 61.4 mbar, 1K Pot = 1.73 K, 500±100 mbar
The 1K Pot surge happened at about the same temperature as in Run 65. After the surge the Tower temperature fell from 2.09 K to 2.02 K (flat before and after). Looks like some sort of phase transition… Looks like it also happened in Run 64, Run 63, Run 61.
2.17 K is the lambda point of LHe.
1050 Tower = 2.02 K (flat), Fridge(G1) = 59.7 mbar (falling slowly), 1K Pot = 1.73 K / 400±100 mbar
1100 Tower = 2.00 K (flat), Fridge(G1) = 58.0 mbar (falling slowly), 1K Pot = 1.73 K / 400±100 mbar
1110 Tower = 1.95 K (falling), Fridge(G1) = 57.2 mbar (falling more slowly), 1K Pot = 1.73 K / 300±100 mbar
1120 Tower = 1.93 K (flattening), Fridge(G1) = 56.1 mbar (falling slowly), 1K Pot = 1.73 K / 300±50 mbar
1145 Tower = 1.90 K (falling slowly), Fridge(G1) = 53.8 mbar (falling slowly), 1K Pot = 1.72 K / 250±50 mbar
1145 Close V9 leaving 53.4 mbar in dump. Close V3. Turn on 3He pump, which will push a little more mix into condenser. Fridge pressures rise a little then fall again.
1300 Tower = 1.87 K (flattish), Fridge(G1) = 48.3 mbar (falling slowly), 1K Pot = 1.72 K / 175±25 mbar
1305 Start circulation through Trap 2. V6 → 12.8%. Condenser peaks at 130 mbar. 1K Pot pressure spikes off scale then falls. Tower falls to 1.7 K.
1315 V6 → 13.2%. Condenser peaks at 121 mbar. Tower → 1.57 K.
1320 V6 → 13.5%. Condenser peaks at 134 mbar. Tower → 1.48 K.
1324 MC(H35)/Coldplate/Still = 1.42/2.16/1.75 K
1335 V6 → 13.8%. Condenser peaks at 143 mbar. Still → 15 mbar. Tower → 1.29 K. (Cools to about 1.1 K then flattens abruptly)
1345 V6 → 14.2%. Condenser rises to 137 mbar and flattens. Still → 3.8 mbar. Tower → 1.12 K (kinda flat).
1349 MC(H35)/Coldplate/Still = 920 mK/875 mK/1.26 K
1400 V6 → 15.0%. Condenser peaks at 145 mbar. Still → 1.6 mbar. Tower → 590 mK (and falling).
1405 1K Pot = 1.71 K / 150±20 mbar
1410 Condenser has fallen to 136 mbar. Still = 1.1 mbar.
1410 V6 → 100%. Condenser peaks at 142 mbar. Still → 0.19 mbar. Tower → 160 mK (and falling).
1415 MC(H35)/Coldplate/Still = 133 mK/270 mK/1.31 K
1420 Tower = 100 mK
1422 He recovery meter = 111200 cuft
1425 Tower/MC/Coldplate = 65/51/163 mK, Still = 1.24 K / 0.150 mbar, G1/G2 = 113/105 mbar, 1K Pot = 1.72 K / 150±20 mbar.
1440 Tower/MC/Coldplate = 44.0/37.6/161 mK, Still = 1.32 K / 0.120 mbar, G1/G2 = 92/86 mbar, 1K Pot = 1.71 K / 150±20 mbar.
1445 Open V14 to let the rest of the mixture from the trap into circulation. Condenser peaks at 128 mbar, Still peaks at 150 mbar, MC rises to 99 mK then falls again.
1500 Tower/MC/Coldplate = 55.4/46.0/163 mK, Still = 1.28 K / 1.8 mbar (V14 open), G1/G2 = 80/76 mbar, 1K Pot = 1.71 K / 150±20 mbar.
1515 Tower/MC/Coldplate = 41.1/31.6/157 mK, Still = 1.25 K / 0.55 mbar (V14 open), G1/G2 = 87/81 mbar, 1K Pot = 1.71 K / 135±15 mbar.
1520 Still = 0.50 mbar. Close V14.
1530 Tower/MC/Coldplate = 38.8/29.4/161 mK, Still = 1.4 K / 0.120 mbar, G1/G2 = 88/82 mbar, 1K Pot = 1.71 K / 135±15 mbar.
1618 Tower/MC/Coldplate = 37.7/28.6/154 mK, Still = 1.26 K / 0.120 mbar, G1/G2 = 89/84 mbar, 1K Pot = 1.71 K / 130±20 mbar.
1648 Fill LN2 cold trap full.

Next Day