2017/09/22 (Friday)

Previous Day

Status: Successful 98K throughput, initial LHe transfer yesterday.
1055 MC(Tower)/Sorb/1K Pot = 4.70/8.3/6.41 K, Fridge(G1) = 587 mbar, MB = 70.7%.
1055 Turbo = 3.0 nbar, IVC = 30 nbar. Stop pumping on IVC.
1100 Fill LN2 cold trap bath
1100 Connect 4He pump to 1K Pot pumping line, turn on pump, open the two valves connecting pump exhaust to recovery
1130 Electrical checkout at 4K, specifically cross-channel resistances:
Cross-channel continuity through qBias leads (all in nS)
measurements at 92K / 4K
DIB1qI (qA) DIB4qI (qB) DIB1qO (qC) DIB3qO (qD) DIB4qO (qE)
DIB1qI (qA) - - - - -
DIB4qI (qB) 13 / 12.90 - - - -
DIB1qO (qC) 1.1 / 0.07 0.7 / 0.06 - - -
DIB3qO (qD) 7 / 5.40 7 / 5.44 1.0 / 0.07 - -
DIB4qO (qE) 13 / 12.96 13 / 13.00 0.7 / 0.06 7 / 5.44 -
note qObias is pin 11, qIbias is pin 15 on I/O cable

This reinforces the notion that qA, qB, and qE are all shorted to one another.

It also suggests that the other two channels are connected and not shorted, since their conductivities have decreased with cooling, by about the same ratio.

1135 Pump out P2 and G3 volumes, then isolate pump and open 1K Pot speedivalve. Mechanical gauge shows about 0.23 bar. 1K Pot temperature dips from 6.4 K to 4.7 K then rises back to >6 K.
1140 Open valve to 4He pump to start pumping on 1K Pot, needle valve closed. Another temporary dip in 1K Pot temperature.
1145 1K Pot needle valve closed. 1K Pot → 6.5 K / 0.04 mbar
Open 1K Pot needle valve about 10 degrees. 1K Pot → 3.3 K / 4.5 mbar
Open needle valve a few more degrees. 1K Pot → 2.8 K / 10.0 mbar
Open needle valve a few more degrees. 1K Pot → 2.4 K / 24.0 mbar. MC temperature falls below 4 K.
Open needle valve a few more degrees. 1K Pot → 2.1 K / 36.0 mbar.
1200 Tower = 3.33 K, 1K Pot = 2.10 K / 36 mbar, Fridge(G1) = 430 mbar
1210 Tower = 2.95 K, 1K Pot = 1.71 K / 34 mbar, Fridge(G1) = 280 mbar
1220 Tower = 2.37 K, 1K Pot = 1.62 K / 24 mbar, Fridge(G1) = 112 mbar
1230 Tower = 2.12 K, 1K Pot = 1.56 K / 11 mbar, Fridge(G1) = 65.4 mbar
1233 Open needle valve a few more degrees. Wait a few minutes to settle, Tower = 2.07 K, 1K Pot = 1.81 K / 1 in Hg (which is about 33 mbar), Fridge(G1) = 61.5 mbar
1251 Close needle vale a few degrees (equal to previous open tweak). Tower = 2.03 K, 1K Pot = 1.68 K / 75 mbar, Fridge(G1) = 60.1 mbar
1310 Tower = 1.99 K, 1K Pot = 1.70 K / 85 mbar, Fridge(G1) = 57.6 mbar
1405 Tower = 1.90 K, 1K Pot = 1.69 K / 70 mbar, Fridge(G1) = 48.7 mbar
1430 Tower = 1.89 K, 1K Pot = 1.69 K / 70 mbar, Fridge(G1) = 47.4 mbar
1435 Close V9 leaving 48.6 mbar in dump. Close V3. Turn on 3He pump, which will push a little more mix into condenser. Fridge pressures rise a little then fall again.
1435 MC(H35)/Coldplate/Still = 1.91/2.97/2.19 K
1440 Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 160 mbar. 1K Pot pressure spike. MC falls to 1.65 K.
1445 V6 → 13.2%. Condenser peaks at 126 mbar. MC → 1.51 K.
1435 Tower/Coldplate/Still = 1.50/2.14/1.66 K
1450 V6 → 13.5%. Condenser peaks at 133 mbar. Still → 55 mbar. MC → 1.55 K.
1500 V6 → 13.8%. Condenser peaks at 130 mbar. Still → 5 mbar. MC → 1.25 K.
1505 1K Pot = 1.69K, 60 mbar. Pretty steady.
1505 Tower/Coldplate/Still = 1.10/1.28/1.23 K. MC(H35) = 1.16 K.
1510 V6 → 14.2%. Condenser flattish at 133 mbar. Still → 3.6 mbar. MC → 0.92 K.
1515 V6 → 15.0%. Condenser → 136 mbar. Still → 1.6 mbar. MC → 0.81 K.
1515 Tower/Coldplate/Still = 0.73/0.72/1.14 K. MC(H35) = 1.16 K.
1515 V6 → 100%. Condenser peaks at 134 mbar. Still → 0.22 mbar. MC → 0.31 K.
1515 Tower/Coldplate/Still = 0.32/0.52/1.18 K. MC(H35) = 1.16 K.
1540 Condenser = 103 mbar, Still = 0.16 mbar, MC(H35)/Tower/Coldplate = 77/140/312 mK
1545 Condenser = 86 mbar, Still = 0.13 mbar, MC(H35)/Tower/Coldplate = 56/110/284 mK. MC looks like it's flattening out.
1615 Condenser = 83 mbar, Still = 0.13 mbar, MC(H35)/Tower/Coldplate = 52/71/293 mK.
1620 Open V14 to let the rest of the mixture from the trap into circulation. Condenser peaks at 115 mbar, Still peaks at 60 mbar, Tower rises to 111 mK.
1625 Condenser has settled down to 76 mbar, Still 4 mbar and falling exponentially, Tower = 78 mK and falling.
1640 Condenser = 81 mbar, Still = 0.85 mbar, Tower = 70 mK. Close V14.
1645 Fill LN2 cold trap bath
1645 Tower/MC/Coldplate = 68.9/52/308 mK, Still = 1.16 K / 0.130 mbar, G1/G2 = 80/90 mbar, 1K Pot = 1.69 K / 60 mbar. MB = 61.0%

Next Day